Operating Instructions Teledyne API – T101 Operation Manual
90
To select manual output calibration for a particular channel, activate the ANALOG I/O
CONFIGURATION MENU (see Section 4.6.3), then press:
DIAG AIO AOUT AUTO CAL: ON
ON ENTR EXIT
Exit to return to
th e ma in
sample display
DIAG ANALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO AOUTS CA LIBRATED: NO
< SET SET> CAL EX IT
DIAG AIO CONC_OUT_2 RANGE: 5V
SET> EDIT EX IT
Toggles the
auto cal mode
ON/ OFF f or
this analog
output channe l
only.
DIAG AIO CONC_OUT_2:5V, CAL
< SET SET> EDIT EXIT
Press SET> to select the analog output ch an nel to
be config ured. The n press EDIT to continue
ENTR accepts the new setting
and returns to the previ ous
menu. EXIT ign or es the new
setting and returns to the
pre vious m enu.
DIAG AIO CONC_OUT_2 REC OFS: 0 mV
< SET SET> EDIT EXIT
DIAG AIO CONC_OUT_2 AUTO CAL: ON
< SET SET> EDIT EXIT
Now the analog output channels should either be automatically calibrated or they should
be set to manual calibration, which is described next.
4.6.3.3. Manual Analog Output Calibration and Voltage Adjustment
For highest accuracy, the voltages of the analog outputs can be manually calibrated.
Calibration is done through the instrument software with a voltmeter connected across the
output terminals (Figure 4-4). Adjustments are made using the front panel buttons by
setting the zero-point first and then the s
pan-point (Table 4-13).
The software allows this adjustment to b
e made in 100, 10 or 1 count increments.
Table 4-13. Voltage Tolerances for Analog Output Calibration
Full Scale Zero Tolerance Span Voltage Span Tolerance
0.1 VDC ±0.0005V 90 mV ±0.001V
1 VDC ±0.001V 900 mV ±0.001V
5 VDC ±0.002V 4500 mV ±0.003V
10 VDC ±0.004V 4500 mV ±0.006V
NOTE
Outputs configured for 0.1V full scale should always be calibrated manually.