Teledyne T101 Manual de usuario

Busca en linea o descarga Manual de usuario para De medición, control y control Teledyne T101. Teledyne T101 User's Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 371
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
OPERATION MANUAL
MODEL T101
UV FLUORESCENCE H
2
S ANALYZER
Also supports operation of:
Model T102 Analyzer
(when used in conjunction with T102 Addendum, PN 07267)
© TELEDYNE ADVANCED POLLUTION INSTRUMENTATION
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
USA
Toll-free Phone:
800-324-5190
Phone:
858-657-9800
Fax:
858-657-9816
Email:
api-sales@teledyne.com
Website:
http://www.teledyne-api.com/
Copyright 2011 - 2012 07266B DCN6845
Teledyne Advanced Pollution Instrumentation 08 June 2012
Vista de pagina 0
1 2 3 4 5 6 ... 370 371

Indice de contenidos

Pagina 1 - S ANALYZER

OPERATION MANUAL MODEL T101 UV FLUORESCENCE H2S ANALYZER Also supports operation of: Model T102 Analyzer (when used in conjunction with T102 A

Pagina 2

viii This page intentionally left blank. 07266B DCN6485

Pagina 3

Operating Instructions Teledyne API – T101 Operation Manual 98 4.6.5. ELECTRICAL TEST The electrical test function creates a current, which substi

Pagina 4

Teledyne API – T101 Operation Manual Operating Instructions 99 4.6.6. LAMP CALIBRATION An important factor in accurately determining H2S concentrati

Pagina 5 - SAFETY MESSAGES

Operating Instructions Teledyne API – T101 Operation Manual 100 4.6.7. PRESSURE CALIBRATION A sensor at the exit of the sample chamber continuousl

Pagina 6 - CONSIGNES DE SÉCURITÉ

Teledyne API – T101 Operation Manual Operating Instructions 101 4.6.8. FLOW CALIBRATION The flow calibration allows the user to adjust the values of

Pagina 7 - WARRANTY

Operating Instructions Teledyne API – T101 Operation Manual 102 4.6.9. TEST CHANNEL OUTPUT When activated, output channel A4 can be used to report

Pagina 8

Teledyne API – T101 Operation Manual Operating Instructions 103 Once a TEST function is selected, the instrument begins to report a signal on the A4

Pagina 9 - REVISION HISTORY

Operating Instructions Teledyne API – T101 Operation Manual 104 To edit the instrument’s ID code, press: SETUP X.X COMMUNICATIONS MENU

Pagina 10 - 07266B DCN6485

Teledyne API – T101 Operation Manual Operating Instructions 105 4.7.2. COM PORT DEFAULT SETTINGS As received from the factory, the analyzer is set u

Pagina 11 - TABLE OF CONTENTS

Operating Instructions Teledyne API – T101 Operation Manual 106 The signals from these two connectors are routed from the motherboard via a wiring

Pagina 12

Teledyne API – T101 Operation Manual Operating Instructions 107 4.7.4. RS-485 CONFIGURATION As delivered from the factory, COM2 is configured for R

Pagina 13

ix TABLE OF CONTENTS 1. INTRODUCTION ...

Pagina 14

Operating Instructions Teledyne API – T101 Operation Manual 108 will automatically be assigned an IP address by the DHCP server (Section 4.10.6.2)

Pagina 15

Teledyne API – T101 Operation Manual Operating Instructions 109 4.7.6.2. Manually Configuring the Ethernet with Static IP Addresses 1. Connect a c

Pagina 16 - LIST OF APPENDICES

Operating Instructions Teledyne API – T101 Operation Manual 110 ENTR accept new settings EXIT ignores new settings SETUP X.X PRIMARY

Pagina 17 - LIST OF FIGURES

Teledyne API – T101 Operation Manual Operating Instructions 111 Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK addresses: Internet Confi

Pagina 18 - LIST OF TABLES

Operating Instructions Teledyne API – T101 Operation Manual 112 4.7.6.3. Changing the Analyzer’s HOSTNAME The HOSTNAME is the name by which the

Pagina 19

Teledyne API – T101 Operation Manual Operating Instructions 113 Table 4-18. Internet Configuration Touchscreen Button Functions BUTTON FUNCTION <

Pagina 20

Operating Instructions Teledyne API – T101 Operation Manual 114 4.7.7. USB CONFIGURATION The analyzer can be operated through a personal computer

Pagina 21 - 1. INTRODUCTION

Teledyne API – T101 Operation Manual Operating Instructions 115 5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following setti

Pagina 22 - 1.2. OPTIONS

Operating Instructions Teledyne API – T101 Operation Manual 116 After connecting a USB cable between your PC and the instrument, ensure their baud

Pagina 23

Teledyne API – T101 Operation Manual Operating Instructions 117 Figure 4-9. Jumper and Cables for Multidrop Mode Note If you are adding an instru

Pagina 24

TABLE OF CONTENTS Teledyne API – T101 Operation Manual x 4.4.3. Setup – DAS: Data Acquisition ...

Pagina 25 - DESCRIPTION

Operating Instructions Teledyne API – T101 Operation Manual 118 Analyzer Analyzer Analyzer Last Analyzer Female DB9Male DB9 RS-232 COM2 RS-232

Pagina 26 - 2.2.2. EMC

Teledyne API – T101 Operation Manual Operating Instructions 119 4.7.9. MODBUS SET UP The following set of instructions assumes that the user is fami

Pagina 27 - 3. GETTING STARTED

Operating Instructions Teledyne API – T101 Operation Manual 120 Example Read/Write Definition window: Example Connection Setup window: Exampl

Pagina 28 - 3.2. INSTRUMENT LAYOUT

Teledyne API – T101 Operation Manual Operating Instructions 121 4.7.10. COM PORT COMMUNICATION MODES Each of the analyzer’s serial ports can be conf

Pagina 29

Operating Instructions Teledyne API – T101 Operation Manual 122 Press the following buttons to select a communication mode for a one of the COMM

Pagina 30

Teledyne API – T101 Operation Manual Operating Instructions 123 4.7.11. COM PORT BAUD RATE To select the baud rate of one of the COM Ports, press:

Pagina 31

Operating Instructions Teledyne API – T101 Operation Manual 124 4.7.12. COM PORT TESTING The serial ports can be tested for correct connection and

Pagina 32 - 3.2.2. REAR PANEL

Teledyne API – T101 Operation Manual Operating Instructions 125 the DAS (Section 4.8). The APICOM manual, which is included with the program, contai

Pagina 33

Operating Instructions Teledyne API – T101 Operation Manual 126 Table 4-21. DAS Data Channel Properties PROPERTY DESCRIPTION DEFAULT SETTING RANGE

Pagina 34

Teledyne API – T101 Operation Manual Operating Instructions 127 Table 4-22. DAS Data Parameter Functions FUNCTION EFFECT PARAMETER Instrument-speci

Pagina 35 - 3.3. ELECTRICAL CONNECTIONS

Teledyne API – T101 Operation Manual TABLE OF CONTENTS xi 4.8.2.9. Starting Date ...

Pagina 36 - ANALOG OUT

Operating Instructions Teledyne API – T101 Operation Manual 128 4.8.2. DEFAULT DAS CHANNELS A set of default Data Channels has been included in th

Pagina 37

Teledyne API – T101 Operation Manual Operating Instructions 129 LIST OF PARAMETERS NAME: PNUMTC EVENT: ATIMER REPORT PERIOD: 001:00:00 NO. OF RECOR

Pagina 38

Operating Instructions Teledyne API – T101 Operation Manual 130 4.8.2.1. Viewing DAS Data and Settings DAS data and settings can be viewed on the

Pagina 39

Teledyne API – T101 Operation Manual Operating Instructions 131 4.8.2.2. Editing DAS Data Channels DAS configuration is most conveniently done throu

Pagina 40

Operating Instructions Teledyne API – T101 Operation Manual 132 PARAMETERS: Four parameters are included in this channel EVENT: This channel is se

Pagina 41 - 3.4. PNEUMATIC CONNECTIONS

Teledyne API – T101 Operation Manual Operating Instructions 133 4.8.2.3. Trigger Events To edit the list of data parameters associated with a specif

Pagina 42

Operating Instructions Teledyne API – T101 Operation Manual 134 4.8.2.4. Editing DAS Parameters Data channels can be edited individually from the

Pagina 43

Teledyne API – T101 Operation Manual Operating Instructions 135 To configure the parameters for a specific data parameter, press: FROM THE EDIT DA

Pagina 44 - CO NTR O L

Operating Instructions Teledyne API – T101 Operation Manual 136 SAMPLE PERIOD is only used when the DAS parameter’s sample mode is set for AVG, MI

Pagina 45

Teledyne API – T101 Operation Manual Operating Instructions 137 To define the REPORT PERIOD, follow the instruction shown in Section 4.11.2.2 then p

Pagina 46

TABLE OF CONTENTS Teledyne API – T101 Operation Manual xii 7.5. Dynamic Multipoint Span Calibration ...

Pagina 47

Operating Instructions Teledyne API – T101 Operation Manual 138 combined number of records would be exceeded. (For information regarding the maxim

Pagina 48

Teledyne API – T101 Operation Manual Operating Instructions 139 4.8.2.7. RS-232 Report Function The T101 DAS can automatically report data to the co

Pagina 49

Operating Instructions Teledyne API – T101 Operation Manual 140 4.8.2.10. Disabling/Enabling Data Channels Data channels can be temporarily disabl

Pagina 50 - Permeation Tube Heater

Teledyne API – T101 Operation Manual Operating Instructions 141 4.8.2.11. HOLDOFF Feature The DAS HOLDOFF feature allows to prevent data collection

Pagina 51 - NO

Operating Instructions Teledyne API – T101 Operation Manual 142 4.8.3. REMOTE DAS CONFIGURATION Editing channels, parameters and triggering events

Pagina 52 - CALIBRATION

Teledyne API – T101 Operation Manual Remote Operation 143 5. REMOTE OPERATION 5.1.1. REMOTE OPERATION USING THE EXTERNAL DIGITAL I/O 5.1.1.1. Stat

Pagina 53

Remote Operation Teledyne API – T101 Operation Manual 144 Connect to Internal Ground of MonitoringSTATUS 1 2 3 4 5 6

Pagina 54 - 3.5.4. FUNCTIONAL CHECK

Teledyne API – T101 Operation Manual Remote Operation 145 Table 5-1. Status Output Pin Assignments CONNECTOR PIN STATUS CONDITION (ON=CONDUCTING)

Pagina 55 - 3.6. INITIAL CALIBRATION

Remote Operation Teledyne API – T101 Operation Manual 146 There are two methods to activate control inputs. The internal +5V available from the “+

Pagina 56

Teledyne API – T101 Operation Manual Remote Operation 147 5.1.2.2. Help Commands in Terminal Mode Table 5-3. Terminal Mode Software Commands COMMAN

Pagina 57

Teledyne API – T101 Operation Manual TABLE OF CONTENTS xiii 9.5.12. PMT Sensor ...

Pagina 58 - S MEASUREMENTS

Remote Operation Teledyne API – T101 Operation Manual 148 5.1.2.3. Command Syntax Commands are not case-sensitive and all arguments within one com

Pagina 59 - 4. OPERATING INSTRUCTIONS

Teledyne API – T101 Operation Manual Remote Operation 149 hexadecimal digits (0-9, A-F, a-f), which is the ‘C’ programming language convention. No

Pagina 60 - 4.2. SAMPLE MODE

Remote Operation Teledyne API – T101 Operation Manual 150 5.1.2.6. General Message Format All messages from the instrument (including those in res

Pagina 61

Teledyne API – T101 Operation Manual Remote Operation 151 To change this setting press: SETUP X.X SECONDARY SETUP MENU COMM VARS DIAG

Pagina 62

Remote Operation Teledyne API – T101 Operation Manual 152 To Initialize the modem press: Select which COM Port is tested SETUP X.X SECONDA

Pagina 63 - 4.2.2. WARNING MESSAGES

Teledyne API – T101 Operation Manual Remote Operation 153 5.1.2.8. COM Port Password Security In order to provide security for remote access of the

Pagina 64 - 4.3. CALIBRATION MODE

Remote Operation Teledyne API – T101 Operation Manual 154 Figure 5-4. APICOM Remote Control Program Interface APICOM is included free of cost wit

Pagina 65

Teledyne API – T101 Operation Manual Remote Operation 155 5.1.4. USING THE T101 WITH A HESSEN PROTOCOL NETWORK 5.1.4.1. General Overview of Hessen

Pagina 66 - 4.4. SETUP MODE

Remote Operation Teledyne API – T101 Operation Manual 156 5.1.4.3. Activating Hessen Protocol The first step in configuring the T101 to operate ov

Pagina 67

Teledyne API – T101 Operation Manual Remote Operation 157 5.1.4.4. Selecting a Hessen Protocol Type Currently there are two version of Hessen Proto

Pagina 68

TABLE OF CONTENTS Teledyne API – T101 Operation Manual xiv 10.4.10.1. A to D Conversion ...

Pagina 69

Remote Operation Teledyne API – T101 Operation Manual 158 To Select a Hessen response mode, press: SETUP X.X SECONDARY SETUP MENU COMM VA

Pagina 70

Teledyne API – T101 Operation Manual Remote Operation 159 5.1.4.6. Hessen Protocol Gas ID Since the T101 can be, when the proper optional equipment

Pagina 71

Remote Operation Teledyne API – T101 Operation Manual 160 5.1.4.7. Setting Hessen Protocol Status Flags Teledyne API’s implementation of Hessen pr

Pagina 72

Teledyne API – T101 Operation Manual Remote Operation 161 To assign or reset the status flag bit assignments, press: SETUP X.X SECONDARY SETUP

Pagina 73

Remote Operation Teledyne API – T101 Operation Manual 162 This page intentionally left blank. 07266B DCN6485

Pagina 74 - 4.4.4.7. Range Units

Teledyne API – T101 Operation Manual Calibration Procedures 163 6. CALIBRATION PROCEDURES This Section describes the calibration procedures for the

Pagina 75 - 4.4.4.8. Dilution Ratio

Calibration Procedures Teledyne API – T101 Operation Manual 164 6.1.2. ZERO AIR Zero air is similar in chemical composition to the Earth’s atmosph

Pagina 76 - Table 4-6. Password Levels

Teledyne API – T101 Operation Manual Calibration Procedures 165 6.1.5. CALIBRATION GAS TRACEABILITY All equipment used to produce calibration gases

Pagina 77 - PRIMARY SETUP MENU

Calibration Procedures Teledyne API – T101 Operation Manual 166 STEP ONE: Connect the sources of zero air and span gas as shown below. No Valve O

Pagina 78 - SECONDARY SETUP MENU

Teledyne API – T101 Operation Manual Calibration Procedures 167 STEP TWO: Set the expected H2S span gas concentrations. In this example the inst

Pagina 79

Teledyne API – T101 Operation Manual TABLE OF CONTENTS xv LIST OF FIGURES Figure 3-1. Front Panel Layout ...

Pagina 80 - ALLOWED

Calibration Procedures Teledyne API – T101 Operation Manual 168 STEP THREE: Perform the zero/span calibration: M-P CAL H2S STB=X.XX

Pagina 81 - See Section 6.8.1

Teledyne API – T101 Operation Manual Calibration Procedures 169 6.3. MANUAL CALIBRATION CHECKS Informal calibration checks, which only evaluate but

Pagina 82 - Press the PREV

Calibration Procedures Teledyne API – T101 Operation Manual 170 6.4. MANUAL CALIBRATION WITH ZERO/SPAN VALVES Zero and Span calibrations using the

Pagina 83 - FUNCTIONS

Teledyne API – T101 Operation Manual Calibration Procedures 171 Step Two: Set the expected H2S span gas value: This seque nce ca uses the analyze

Pagina 84

Calibration Procedures Teledyne API – T101 Operation Manual 172 Step Three: Perform the calibration or calibration check according to the followin

Pagina 85 - 4.6.1. SIGNAL I/O

Teledyne API – T101 Operation Manual Calibration Procedures 173 6.5. MANUAL CALIBRATION WITH IZS OPTION The following section describes the basic m

Pagina 86

Calibration Procedures Teledyne API – T101 Operation Manual 174 6.6. MANUAL CALIBRATION CHECKS WITH IZS OR ZERO/SPAN VALVES Zero and span checks u

Pagina 87

Teledyne API – T101 Operation Manual Calibration Procedures 175 STEP ONE: Connect the sources of Zero Air and Span Gas as shown below. Exter

Pagina 88 - Table 4-12

Calibration Procedures Teledyne API – T101 Operation Manual 176 STEP TWO: Perform the zero/span check. Wait until H2S STB falls below 0.5 pp b.

Pagina 89

Teledyne API – T101 Operation Manual Calibration Procedures 177 6.7. MANUAL CALIBRATION IN INDEPENDENT OR AUTO REPORTING RANGE MODES If the analyze

Pagina 90

TABLE OF CONTENTS Teledyne API – T101 Operation Manual xvi Figure 10-8. Typical Flow Control Assembly with Critical Flow Orifice ...

Pagina 91

Calibration Procedures Teledyne API – T101 Operation Manual 178 If contact closures are used in conjunction with the analyzer’s AutoCal (Section 6

Pagina 92

Teledyne API – T101 Operation Manual Calibration Procedures 179 The analyzer will ask to have the GAS TYPE specified at the beginning of the proces

Pagina 93 - 4.6.3), then press:

Calibration Procedures Teledyne API – T101 Operation Manual 180 Table 6-3. AutoCal Attribute Setup Parameters Attribute Name ACTION Timer Enabled

Pagina 94

Teledyne API – T101 Operation Manual Calibration Procedures 181 The following example sets Sequence2 to carry out a zero-span calibration every oth

Pagina 95 - (Figure 4-5)

Calibration Procedures Teledyne API – T101 Operation Manual 182 Toggle to se t number of days between procedures (1-367) Toggle to set delay time

Pagina 96

Teledyne API – T101 Operation Manual Calibration Procedures 183 It is recommended that, if dynamic calibration is enabled, the analyzer’s test func

Pagina 97 - 4.6.3.6. AIN Calibration

Calibration Procedures Teledyne API – T101 Operation Manual 184 6.9.2. AUTOCAL OF INSTRUMENTS IN MULTIGAS MEASUREMENT MODE If the analyzer is bein

Pagina 98

Teledyne API – T101 Operation Manual Calibration Procedures 185 6.10. CALIBRATION QUALITY After completing one of the calibration procedures descri

Pagina 99 - 4.6.4. OPTIC TEST

Calibration Procedures Teledyne API – T101 Operation Manual 186 This page intentionally left blank. 07266B DCN6485

Pagina 100 - 4.6.5. ELECTRICAL TEST

187 7. EPA PROTOCOL CALIBRATION 7.1. CALIBRATION REQUIREMENTS When configured for SO2 measurement mode or multigas measurement mode, the T101 may b

Pagina 101 - 4.6.6. LAMP CALIBRATION

Teledyne API – T101 Operation Manual TABLE OF CONTENTS xvii Table 4-14. Current Loop Output Calibration with Resistor ...

Pagina 102 - 4.6.7. PRESSURE CALIBRATION

EPA Protocol Calibration Teledyne API – T101 Operation Manual 188 time. Therefore, it is necessary to dynamically check the calibration relationsh

Pagina 103 - 4.6.8. FLOW CALIBRATION

Teledyne API – T101 Operation Manual EPA Protocol Calibration 189 Table 7-2. Activity Matrix for Calibration Procedure EQUIPMENT/SUPPLIES ACCEPTANC

Pagina 104 - 4.6.9. TEST CHANNEL OUTPUT

EPA Protocol Calibration Teledyne API – T101 Operation Manual 190 7.1.5. CALIBRATION FREQUENCY To ensure accurate measurements of the SO2 concentr

Pagina 105 - COMMUNICATION PORTS

Teledyne API – T101 Operation Manual EPA Protocol Calibration 191 7.1.7. SUMMARY OF QUALITY ASSURANCE CHECKS The following items should be checked

Pagina 106

EPA Protocol Calibration Teledyne API – T101 Operation Manual 192 Table 7-4. Definition of Level 1 and Level 2 Zero and Span Checks (from Section

Pagina 107

Teledyne API – T101 Operation Manual EPA Protocol Calibration 193 7.3. ZERO AND SPAN CHECKS A system of Level 1 and Level 2 zero span checks (see

Pagina 108

EPA Protocol Calibration Teledyne API – T101 Operation Manual 194  Single range selection. See Section 4.7.4 of this manual. If the instrument w

Pagina 109 - 4.7.4. RS-485 CONFIGURATION

Teledyne API – T101 Operation Manual EPA Protocol Calibration 195 7.5. DYNAMIC MULTIPOINT SPAN CALIBRATION Dynamic calibration involves introducing

Pagina 110

EPA Protocol Calibration Teledyne API – T101 Operation Manual 196 7.6. SPECIAL CALIBRATION REQUIREMENTS FOR INDEPENDENT RANGE OR AUTO RANGE If Ind

Pagina 111

197 8. INSTRUMENT MAINTENANCE Predictive diagnostic functions including data acquisition, failure warnings and alarms built into the analyzer allow

Pagina 113 - Network Administrator

TABLE OF CONTENTS Teledyne API – T101 Operation Manual xviii This page intentionally left blank. 07266B DCN6485

Pagina 114

Instrument Maintenance Teledyne API – T101 Operation Manual 198 047400102 Rev XC This page intentionally left blank. 07266B DCN6485

Pagina 115 - >\

Instrument Maintenance Model T101 Instruction Manual 199 Table 8-1 T101 Preventive Maintenance Schedule ITEM ACTION FREQUENCY CAL CHECK MANUAL SE

Pagina 116 - 4.7.7. USB CONFIGURATION

Instrument Maintenance Model T101 Instruction Manual 200 This page intentionally left blank. 07266B DCN6485

Pagina 117

Model T101 Instruction Manual Instrument Maintenance 201 8.2. PREDICTIVE DIAGNOSTICS The analyzer’s test functions can be used to predict failure

Pagina 118

Instrument Maintenance Model T101 Instruction Manual 202 8.3. MAINTENANCE PROCEDURES The following procedures need to be performed regularly as part

Pagina 119

Model T101 Instruction Manual Instrument Maintenance 203 8.3.2. CHANGING THE IZS PERMEATION TUBE 1. Turn off the analyzer, unplug the power cord a

Pagina 120 - RS-232 port

Instrument Maintenance Model T101 Instruction Manual 204 EXAMPLE: If the SO2 concentration is 125 ppb: Operational hours = 1000 ppm/hr ÷ 0.125 pp

Pagina 121 - 4.7.9. MODBUS SET UP

Model T101 Instruction Manual Instrument Maintenance 205 8.3.4. CHANGING THE EXTERNAL ZERO AIR SCRUBBER The chemicals in the external scrubber need

Pagina 122

Instrument Maintenance Model T101 Instruction Manual 206 8.3.5. MAINTAINING THE H2S  SO2 CONVERTER The catalyst contained in the H2S  SO2 converte

Pagina 123 - ID DESCRIPTION

Model T101 Instruction Manual Instrument Maintenance 207 Converter Efficiency = 490.3 ÷ 499.1 Converter Efficiency = 0.982 (98.2%) 10. It is

Pagina 124

19 1. INTRODUCTION The Model T101 UV Fluorescence H2S Analyzer measures hydrogen sulfide in levels commonly required for Ambient Air monitoring. Th

Pagina 125 - 4.7.11. COM PORT BAUD RATE

Instrument Maintenance Model T101 Instruction Manual 208 Converter Assembly Cover Band Heater and T/C Assembly H2S  SO2 converter Converter Assembly

Pagina 126 - 4.7.12. COM PORT TESTING

Model T101 Instruction Manual Instrument Maintenance 209 8.3.6. CHECKING FOR LIGHT LEAKS When re-assembled or operated improperly, the T101 can dev

Pagina 127 - 4.8.1. DAS STRUCTURE

Instrument Maintenance Model T101 Instruction Manual 210 Gas Line fitting Spring Sintered Filter O-Ring Critical Flow Orifice O-Ring Vacuum Manifold

Pagina 128 - 4.8.1.2. DAS Parameters

Model T101 Instruction Manual Troubleshooting & Service 211 9. TROUBLESHOOTING & SERVICE This section contains a variety of methods for i

Pagina 129

Troubleshooting & Service Model T101 Instruction Manual 212  Examine the values of all TEST functions and compare them to factory values. Note

Pagina 130 - 4.8.2. DEFAULT DAS CHANNELS

Model T101 Instruction Manual Troubleshooting & Service 213 (Fault LED lit, warning msg in Param field, CLR button active) (Fault LED lit, Te

Pagina 131 - LIST OF CHANNELS

Troubleshooting & Service Model T101 Instruction Manual 214 To view or clear a warning message press: SAMPLE SYSTEM RESET H

Pagina 132

Model T101 Instruction Manual Troubleshooting & Service 215 WARNING MESSAGE FAULT CONDITION POSSIBLE CAUSES IZS TEMP WARNING On units with IZS

Pagina 133 - 0) CONC: ATIMER, 4, 800

Troubleshooting & Service Model T101 Instruction Manual 216 9.1.2. FAULT DIAGNOSIS WITH TEST FUNCTIONS Besides being useful as predictive diagnos

Pagina 134

Model T101 Instruction Manual Troubleshooting & Service 217 9.1.3. USING THE DIAGNOSTIC SIGNAL I/O FUNCTION The signal I/O parameters found und

Pagina 135 - 4.8.2.3. Trigger Events

Introduction Teledyne API – T101 Operation Manual 20 1.2. OPTIONS The options available for your analyzer are presented in Table 1-1 with name, op

Pagina 136 - Press SET> until…

Troubleshooting & Service Model T101 Instruction Manual 218 Toggles parameter ON/OFFExit returns to DIAG display & all values return to

Pagina 137

Model T101 Instruction Manual Troubleshooting & Service 219 9.1.4.1. Motherboard Status Indicator (Watchdog) DS5, a red LED on the upper portio

Pagina 138

Troubleshooting & Service Model T101 Instruction Manual 220 Table 9-3. Relay Board Status LEDs LED COLOR FUNCTION FAULT STATUS INDICATED FAILUR

Pagina 139 - 4.8.2.6. Number of Records

Model T101 Instruction Manual Troubleshooting & Service 221 9.2.1. ZERO OR LOW SAMPLE FLOW If the pump is operating but the unit reports a 0 ga

Pagina 140

Troubleshooting & Service Model T101 Instruction Manual 222  If the response offset test function for H2S (H2S OFFS) are greater than 150 mV, a

Pagina 141 - 4.8.2.9. Starting Date

Model T101 Instruction Manual Troubleshooting & Service 223  Verify that the expected concentration is set properly to the actual span gas co

Pagina 142

Troubleshooting & Service Model T101 Instruction Manual 224  Dilution air contains sample or span gas.  Sample inlet may be contaminated with

Pagina 143 - 4.8.2.11. HOLDOFF Feature

Model T101 Instruction Manual Troubleshooting & Service 225  Dirty sample chamber. Clean the sample chamber.  Insufficient time allowed for

Pagina 144

Troubleshooting & Service Model T101 Instruction Manual 226 bubble solution, looking for fine bubbles. Once the fittings have been wetted with so

Pagina 145 - 5. REMOTE OPERATION

Model T101 Instruction Manual Troubleshooting & Service 227 If the unit is set for 220-240 V and is plugged into 100-120 V, the analyzer will n

Pagina 146

Teledyne API – T101 Operation Manual Introduction 21 Option Option Number Description/Notes Analog Inputs Used for connecting external voltage sign

Pagina 147 - 5.1.1.2. Control Inputs

Troubleshooting & Service Model T101 Instruction Manual 228 9.5.5. I2C BUS Operation of the I2C bus can be verified by observing the behavior of

Pagina 148

Model T101 Instruction Manual Troubleshooting & Service 229 Table 9-6. Relay Board Control Devices FUNCTION CONTROL DEVICE SOCKETED Valve0 – Va

Pagina 149

Troubleshooting & Service Model T101 Instruction Manual 230 9.5.9.3. Status Outputs The procedure below can be used to test the Status outputs.

Pagina 150 - 5.1.2.4. Data Types

Model T101 Instruction Manual Troubleshooting & Service 231 In some rare circumstances, this failure may be caused by a bad IC on the motherboa

Pagina 151 - 5.1.2.5. Status Reporting

Troubleshooting & Service Model T101 Instruction Manual 232 Further help with serial communications is available in a separate manual “RS-232 Man

Pagina 152

Model T101 Instruction Manual Troubleshooting & Service 233 9.5.15. HIGH VOLTAGE POWER SUPPLY The HVPS is located in the interior of the sensor

Pagina 153

Troubleshooting & Service Model T101 Instruction Manual 234 The IZS option is heated with a proportional heater circuit and the temperature is ma

Pagina 154

Model T101 Instruction Manual Troubleshooting & Service 235 be re-entered before the instrument will function correctly. Also, zero and span ca

Pagina 155

Troubleshooting & Service Model T101 Instruction Manual 236 Figure 9-4. Shutter Assembly CAUTION ALWAYS wear UV-Protective, Safety Glasses when

Pagina 156

Model T101 Instruction Manual Troubleshooting & Service 237 Table 9-9. Example of UV Lamp Power Supply Outputs UVLAMP_SIGNAL ACTION TO BE TAKE

Pagina 157 - Duplex Full Half

Introduction Teledyne API – T101 Operation Manual 22 Option Option Number Description/Notes RS-232 Multidrop Enables communications between host c

Pagina 158

Troubleshooting & Service Model T101 Instruction Manual 238 NOTE DO NOT grasp the UV lamp by its cap when changing its position (refer to Figure

Pagina 159

Model T101 Instruction Manual Troubleshooting & Service 239 Figure 9-6. Pre-Amplifier Board Layout 6. Turn the gain adjustment potentiometer

Pagina 160

Troubleshooting & Service Model T101 Instruction Manual 240 9.7. FREQUENTLY ASKED QUESTIONS (FAQS) The following list contains some of the most c

Pagina 161

Model T101 Instruction Manual Troubleshooting & Service 241 QUESTION ANSWER How often do I need to change the particulate filter? Once per week

Pagina 162

Troubleshooting & Service Model T101 Instruction Manual 242 This page intentionally left blank. 07266B DCN6485

Pagina 163

Model T101 Instruction Manual Principles Of Operation 243 10. PRINCIPLES OF OPERATION The T101 UV Fluorescence H2S Analyzer is a microprocessor c

Pagina 164

Principles Of Operation Model T101 Instruction Manual 244 The chemical process is: 2222SO2OH2O3SH2  The converter is a heated stainless steel c

Pagina 165 - 6. CALIBRATION PROCEDURES

Model T101 Instruction Manual Principles Of Operation 245 Figure 10-1. UV Absorption The equation for defining the average intensity of the UV lig

Pagina 166 - 6.1.4. PERMEATION TUBES

Principles Of Operation Model T101 Instruction Manual 246 *SOkF2 Where: F = the amount of fluorescent light given off. k = The rate at whic

Pagina 167 - 6.2. MANUAL CALIBRATION

Model T101 Instruction Manual Principles Of Operation 247 10.2. THE UV LIGHT PATH The optical design of the Model T101’s sample chamber optimizes t

Pagina 168 - Chassis

23 2. SPECIFICATIONS AND APPROVALS 2.1. SPECIFICATIONS Table 2-1. Model T101 Basic Unit Specifications PARAMETER DESCRIPTION Ranges H2S: Min 0-50 p

Pagina 169 - STEP TWO: Set the expected H

Principles Of Operation Model T101 Instruction Manual 248 The lamp used in the Model T101 is constructed with a vacuum jacket surrounding a double-bo

Pagina 170

Model T101 Instruction Manual Principles Of Operation 249 10.2.4. OPTICAL FILTERS The Model T101 analyzer uses two stages of optical filters to enh

Pagina 171

Principles Of Operation Model T101 Instruction Manual 250 LAMP OUTPUT (Arbitrary Untis) 1051041021031011100 200 300 400 500 WAVELENGTH (nm) 0213.9 3

Pagina 172 - (At high

Model T101 Instruction Manual Principles Of Operation 251 Another lens located between the excitation UV source lamp and the sample chamber collima

Pagina 173 - span gas value:

Principles Of Operation Model T101 Instruction Manual 252 10.2.6.2. UV Absorption by Ozone Because ozone absorbs UV Light over a relatively broad spe

Pagina 174

Model T101 Instruction Manual Principles Of Operation 253 10.3. PNEUMATIC OPERATION CAUTION It is important that the sample airflow system is leak-

Pagina 175

Principles Of Operation Model T101 Instruction Manual 254 10.3.1. SAMPLE GAS FLOW The flow of gas through the T101 UV Fluorescence H2S Analyzer is cr

Pagina 176 - ZERO/SPAN VALVES

Model T101 Instruction Manual Principles Of Operation 255 10.3.2. MULTIGAS MEASUREMENT & H2S  SO2 SWITCHING VALVE. When activated for operatio

Pagina 177

Principles Of Operation Model T101 Instruction Manual 256 pressure differential combined with the action of the analyzer’s external pump draws the ga

Pagina 178

Model T101 Instruction Manual Principles Of Operation 257 10.3.5. HYDROCARBON SCRUBBER (KICKER) It is very important to make sure the air supplied

Pagina 179

Specifications and Approvals Teledyne API – T101 Operation Manual 24 PARAMETER DESCRIPTION Standard I/O 1 Ethernet: 10/100Base-T 2 RS-232 (300 –

Pagina 180 - No Valve Options Installed

Principles Of Operation Model T101 Instruction Manual 258 10.3.7. PNEUMATIC SENSORS The T101 uses two pneumatic sensors to verify gas streams. These

Pagina 181 - Table 6-2. AutoCal Modes

Model T101 Instruction Manual Principles Of Operation 259 10.4. ELECTRONIC OPERATION Analog In Pneumatic Sensor Board Sample Pressure Sensor Samp

Pagina 182

Principles Of Operation Model T101 Instruction Manual 260 The core of the analyzer is a microcomputer that controls various internal processes, inter

Pagina 183 - 0.5 hours later each day

Model T101 Instruction Manual Principles Of Operation 261 10.4.1. CPU The unit’s CPU card, installed on the motherboard located inside the rear pan

Pagina 184 - Default

Principles Of Operation Model T101 Instruction Manual 262 10.4.2. SENSOR MODULE & SAMPLE CHAMBER Electronically, the T101 sensor module is a grou

Pagina 185 - REPORTING RANGE MODES

Model T101 Instruction Manual Principles Of Operation 263 10.4.4. PHOTO MULTIPLIER TUBE (PMT) The T101 uses a photo multiplier tube (PMT) to detect

Pagina 186 - MEASUREMENT MODE

Principles Of Operation Model T101 Instruction Manual 264 Figure 10-14. Basic PMT Design A significant performance characteristic of the PMT is the

Pagina 187 - 6.10. CALIBRATION QUALITY

Model T101 Instruction Manual Principles Of Operation 265 TEC Control PCA PMT Preamp PCA Thermistor outputs temp of cold block to preamp PCA Pr

Pagina 188

Principles Of Operation Model T101 Instruction Manual 266 be adjusted through a potentiometer. These adjustments should only be performed when encoun

Pagina 189 - 7. EPA PROTOCOL CALIBRATION

Model T101 Instruction Manual Principles Of Operation 267 10.4.7. PNEUMATIC SENSOR BOARD The flow and pressure sensors of the T101 are located on a

Pagina 190

25 3. GETTING STARTED 3.1. UNPACKING AND INITIAL SETUP CAUTION To avoid personal injury, always use two persons to lift and carry the Model T101.

Pagina 191 - TRACEABILITY

Principles Of Operation Model T101 Instruction Manual 268 10.4.9. STATUS LEDS & WATCH DOG CIRCUITRY Zero/Span and IZS Options Sample/Cal Valve I

Pagina 192 - 7.1.6. RECORD KEEPING

Model T101 Instruction Manual Principles Of Operation 269 As a Safety measure, special circuitry on the Relay Board watches the status of LED D1. S

Pagina 193 - Table 7-3. Activity Matrix

Principles Of Operation Model T101 Instruction Manual 270 board. It is digitized and sent to the CPU where it is used to calculate the current temper

Pagina 194

Model T101 Instruction Manual Principles Of Operation 271 In its standard configuration, the analyzer comes with all three of these channels set up

Pagina 195 - 7.3. ZERO AND SPAN CHECKS

Principles Of Operation Model T101 Instruction Manual 272 AC POWER ENTRANCE ON/OFF SWITCH PS 1 (+5 VDC; ±15 VDC) PS 2 (+12 VDC) Chassis Cooling Fa

Pagina 196 - 7.4.2. PRECISION CHECK

Model T101 Instruction Manual Principles Of Operation 273 10.6. FRONT PANEL/DISPLAY INTERFACE Users can input data and receive information directly

Pagina 197

Principles Of Operation Model T101 Instruction Manual 274 10.7. SOFTWARE OPERATION The T101 H2S analyzer is at its heart a high performance, 386-base

Pagina 198 - 7.7. REFERENCES

Model T101 Instruction Manual Principles Of Operation 275 10.7.2. CALIBRATION - SLOPE AND OFFSET Calibration of the analyzer is performed exclusive

Pagina 199 - 8. INSTRUMENT MAINTENANCE

Principles Of Operation Model T101 Instruction Manual 276 10.7.3. TEMPERATURE AND PRESSURE COMPENSATION (TPC) FEATURE As explained previously, change

Pagina 200 - 198 047400102 Rev XC

Model T101 Instruction Manual A Primer on Electro-Static Discharge 277 11. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne API considers the preven

Pagina 201

Getting Started Teledyne API – T101 Operation Manual 26 4. Inspect the interior of the instrument to make sure all circuit boards and other compo

Pagina 202

A Primer on Electro-Static Discharge Model T101 Instruction Manual 278 workbench, using a plastic handled screwdriver or even the constant jostling o

Pagina 203

Model T101 Instruction Manual A Primer on Electro-Static Discharge 279  Any time a charged surface (including the human body) discharges to a dev

Pagina 204 - 8.3. MAINTENANCE PROCEDURES

A Primer on Electro-Static Discharge Model T101 Instruction Manual 280  As long as my analyzer is properly installed, it is safe from damage caused

Pagina 205 - SCRUBBER

Model T101 Instruction Manual A Primer on Electro-Static Discharge 281  Always store sensitive components and assemblies in anti-ESD storage bags

Pagina 206 - To check to see if your SO

A Primer on Electro-Static Discharge Model T101 Instruction Manual 282 11.4.2. BASIC ANTI-ESD PROCEDURES FOR ANALYZER REPAIR AND MAINTENANCE 11.4.2.1

Pagina 207

Model T101 Instruction Manual A Primer on Electro-Static Discharge 283 6. Disconnecting your wrist strap is always the last action taken before le

Pagina 208 - CONVERTER

A Primer on Electro-Static Discharge Model T101 Instruction Manual 284 static charges from damaging the components or assemblies being shipped make s

Pagina 209 - 8.3.5.3. Changing the H

Model T101 Instruction Manual Clossary 285 GLOSSARY Note: Some terms in this glossary may not occur elsewhere in this manual. Term Description/Def

Pagina 210

Glossary Model T101 Instruction Manual 286 Term Description/Definition O2 molecular oxygen O3 ozone SO2 sulfur dioxide cm3 metric abbreviation for c

Pagina 211

Model T101 Instruction Manual Glossary 287 Term Description/Definition FPI Fabry-Perot Interface: a special light filter typically made of a trans

Pagina 212

Teledyne API – T101 Operation Manual Getting Started 27 Figure 3-1. Front Panel Layout Figure 3-2. Display Screen and Touch Control CAUTION –

Pagina 213 - 9.1. GENERAL TROUBLESHOOTING

Glossary Model T101 Instruction Manual 288 Term Description/Definition PFA Per-Fluoro-Alkoxy, an inert polymer; one of the polymers that Du Pont mar

Pagina 214

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-1 APPENDIX A - Version Specific Software Documentation

Pagina 215

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-2 This page intentionally left blank. 07266B DCN6485

Pagina 216

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-3 APPENDIX A-1: Software Menu Trees, S/W Version C.7

Pagina 217

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-4 LOW HIGH ZERO SPAN CONC SAMPLE <TST TST> SET

Pagina 218

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-5 CFG  MODEL NAME  PART NUMBER  SERIAL NUMBER

Pagina 219

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-6 NAME EVENT PARAMETERS REPORT PERIOD NUMBER OF RECOR

Pagina 220 - 9.1.4. STATUS LEDS

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-7 QUIET COMPUTER HESSEN PROTOCOL E, 8,1 E, 7, 1 RS

Pagina 221 - CPU Status LED

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-8 INSTRUMENT IP5 GATEWAY IP5 SUBNET MASK5 TCP PORT3 D

Pagina 222 - 9.2. GAS FLOW PROBLEMS

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-9 . YES NO SAMPLE FLOW WARNING BENCH TEMP WARNI

Pagina 223 - 9.3. CALIBRATION PROBLEMS

i ABOUT TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instrume

Pagina 224 - 9.3.2. NO RESPONSE

Getting Started Teledyne API – T101 Operation Manual 28 The front panel liquid crystal display screen includes touch control. Upon analyzer start-

Pagina 225 - 9.3.6. NON-LINEAR RESPONSE

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-10 37 INTERNAL ANALOG  VOLTAGE SIGNALS 72 (see Ap

Pagina 226 - 9.4.2. SLOW RESPONSE

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-11 APPENDIX A-2: Setup Variables For Serial I/O, S/W Vers

Pagina 227 - 9.5. SUBSYSTEM CHECKOUT

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-12 Setup Variable Numeric Units Default Value Value Ran

Pagina 228

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-13 Setup Variable Numeric Units Default Value Value Ran

Pagina 229 - 9.5.4. DC POWER SUPPLY

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-14 Setup Variable Numeric Units Default Value Value Ran

Pagina 230 - 9.5.7. LCD DISPLAY MODULE

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-15 Setup Variable Numeric Units Default Value Value Ran

Pagina 231 - 9.5.9. MOTHERBOARD

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-16 Setup Variable Numeric Units Default Value Value Ran

Pagina 232 - 9.5.10. CPU

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-17 Setup Variable Numeric Units Default Value Value Ran

Pagina 233 - 9.5.11. RS-232 COMMUNICATION

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-18 Setup Variable Numeric Units Default Value Value Ran

Pagina 234 - 9.5.12. PMT SENSOR

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-19 Setup Variable Numeric Units Default Value Value Ran

Pagina 235 - 9.5.17. IZS OPTION

Teledyne API – T101 Operation Manual Getting Started 29 Figure 3-3. Display/Touch Control Screen Mapped to Menu Charts 07266B DCN6485

Pagina 236 - 9.6. REPAIR PROCEDURES

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-20 Setup Variable Numeric Units Default Value Value Ran

Pagina 237

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-21 APPENDIX A-3: Warnings and Test Measurements, S/W Vers

Pagina 238

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-22 WPMTTEMP PMT TEMP WARNING PMT temperature outside

Pagina 239 - 9.6.3. REPLACING THE UV LAMP

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-23 Table A-3: Test Measurements TEST Measurement M

Pagina 240

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-24 TEST Measurement Message Text DESCRIPTION PMTTEMP

Pagina 241

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-25 APPENDIX A-4: Signal I/O Definitions, S/W Version C.7

Pagina 242

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-26 Signal Name Bit or Channel Number Description 6–7 A

Pagina 243 - 9.8. TECHNICAL ASSISTANCE

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-27 Signal Name Bit or Channel Number Description 1 = al

Pagina 244

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-28 Signal Name Bit or Channel Number Description DARK_S

Pagina 245 - 10. PRINCIPLES OF OPERATION

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-29 Signal Name Bit or Channel Number Description TEMP_I

Pagina 246 - SO2OH2O3SH2 

Getting Started Teledyne API – T101 Operation Manual 30 3.2.2. REAR PANEL Figure 3-4. Rear Panel Layout 07266B DCN6485

Pagina 247 - SOaxexp1IIa

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-30 Table A-5: DAS Trigger Events Name Description

Pagina 248 - = Amount of excited state SO

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-31 Appendix A-6: DAS Functions Name Description Units

Pagina 249 - 10.2.1. UV SOURCE LAMP

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-32 Name Description Units RCTEMP Reaction cell temper

Pagina 250 - 10.2.3. THE PMT

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-33 Name Description Units XIN76 Channel 7 Analog In

Pagina 251 - 10.2.4. OPTICAL FILTERS

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-34 APPENDIX A-7: MODBUS Register Map MODBUS Register

Pagina 252 - 10.2.5. OPTICAL LENSES

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-35 58 Sample pressure “Hg 60 Internal box temperature

Pagina 253

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-36 4 PMT temperature warning 5 Reaction cell temperat

Pagina 254 - 10.2.6.5. Light Pollution

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-37 1 T101, M101E. 2 Concentration alarm option. 3 Set

Pagina 255 - 10.3. PNEUMATIC OPERATION

Teledyne API T/E-Series Models 101, 102, and 108 Software Documentation (PN05492D DCN6485) A-38 This page intentionally left blank. 07266B DCN6485

Pagina 256 - 10.3.1. SAMPLE GAS FLOW

T101 Spare Parts ListPARTNUMBER DESCRIPTION000940100 CD, ORIFICE, .003 GREEN 000940400 CD, ORIFICE, .004 BLUE 000

Pagina 257 - 10.3.3. FLOW RATE CONTROL

Teledyne API – T101 Operation Manual Getting Started 31 Table 3-2. Rear Panel Description Component Function cooling fan Pulls ambient air into cha

Pagina 258

T101 Spare Parts ListPARTNUMBER DESCRIPTION(Reference: 073470000, 1/19/2011 10:10:13 AM) 058021100 PCA, MOTHERBD, GEN 5-ICOP

Pagina 259 - 10.3.6. SO

T101 Spare Parts ListPARTNUMBER DESCRIPTION(Reference: 073470000, 1/19/2011 10:10:13 AM) OR0000094 ORING, 2-228V, 50 DURO VITON(KB)

Pagina 260 - 10.3.7. PNEUMATIC SENSORS

This page intentionally left blank. B-407266B DCN6845

Pagina 261 - CPU Card

Appendix C Warranty/Repair Questionnaire T101, M101E (05494D DCN5798) C-1 Company: _________________________ Contact Name: ______________________

Pagina 262

Appendix C Warranty/Repair Questionnaire T101, M101E (05494D DCN5798) C-2 Notes and further information: _________________________________________

Pagina 263 - 10.4.1. CPU

APPENDIX D – Wire List and Electronic Schematics 07266B DCN6845D-1

Pagina 264

This page intentionally left blank. D-207266B DCN6845

Pagina 265 - Figure 10-13. PMT Assembly

Interconnect List, T101(Reference 0736501 Rev01)RevisionChecked DateDCN01 KV 9/3/10 Cable PN Signal AssemblyPN J/P PinAssemblyPN J/P Pin0337801 CBL A

Pagina 266 - 10.4.5. PMT COOLING SYSTEM

Interconnect List, T101(Reference 0736501 Rev01)Cable PN Signal AssemblyPN J/P PinAssemblyPN J/P PinFROMTO04176CBL, DC POWER TO RELAY BOARDDGND Relay

Pagina 267 - 10.4.6. PMT PREAMPLIFIER

Interconnect List, T101(Reference 0736501 Rev01)Cable PN Signal AssemblyPN J/P PinAssemblyPN J/P PinFROMTO04562 CBL, Z/S IZS VALVESSample Valve +12VRe

Pagina 268 - PMT Preamp PCA

Getting Started Teledyne API – T101 Operation Manual 32 3.2.3. INTERNAL CHASSIS LAYOUT Figure 3-5. Internal Chassis Layout 07266B DCN6485

Pagina 269 - 10.4.8.2. Valve Control

Interconnect List, T101(Reference 0736501 Rev01)Cable PN Signal AssemblyPN J/P PinAssemblyPN J/P PinFROMTO06746CBL, MB TO 06154 CPUGND Motherboard 058

Pagina 270

07266B DCN6845D-7

Pagina 271 - 10.4.10.2. Sensor Inputs

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 30-Jun-2004 Sheet of File: N:\PCBMGR\RELEASED\03954cc\PROTEL\03954a.ddbDrawn By:TeTe03956 A

Pagina 272 - 10.4.11. ANALOG OUTPUTS

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 30-Jun-2004 Sheet of File: N:\PCBMGR\RELEASED\03954cc\PROTEL\03954a.ddbDrawn By:TeTe03956 A

Pagina 273 - 10.4.14. POWER UP CIRCUIT

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 30-Jun-2004 Sheet of File: N:\PCBMGR\RELEASED\03954cc\PROTEL\03954a.ddbDrawn By:TeTe03956 A

Pagina 274

1 2 34ABCD4321DCBAAPPROVALSDRAWNCHECKEDAPPROVEDDATESIZE DRAW ING NO. REVISIONSHEETThe information herein is theproperty of API and issubmitted in stri

Pagina 275

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 10-May-2007 Sheet of File: N:\PCBMGR\04179cc\Source\RevG\04179.ddbDrawn By:H04181M100E/200E

Pagina 276 - 10.7.1. ADAPTIVE FILTER

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 10-May-2007 Sheet of File: N:\PCBMGR\04179cc\Source\RevG\04179.ddbDrawn By:R27 499R8150KR15

Pagina 277

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 10-May-2007 Sheet of File: N:\PCBMGR\04179cc\Source\RevG\04179.ddbDrawn By:R121000MR34.99KR

Pagina 278 - FEATURE

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 10-May-2007 Sheet of File: N:\PCBMGR\04179cc\Source\RevG\04179.ddbDrawn By:48231U16ALF353,

Pagina 279

Teledyne API – T101 Operation Manual Getting Started 33 3.3. ELECTRICAL CONNECTIONS NOTE To maintain compliance with EMC standards, it is required

Pagina 280

1 2 3 4ABCD4321DCBAAPPROVALSDRAWNCHECKEDAPPROVEDDATESIZE DRAWING NO. REVISIONSHEETThe information herein is theproperty of API and issubmitted in stri

Pagina 281

1 2 3 4 5 6ABCD654321DCBAAPPROVALSDRAWNCHECKEDAPPROVEDDATESIZE DRAWING NO. REVISIONSHEETError : LOGO.BMP file not found.The information herein is thep

Pagina 282 - 11.4.1. GENERAL RULES

1 2 3 4 5 6ABCD654321DCBATitleNumber RevisionSizeBDate: 2-Jun-2004 Sheet of File: N:\PCBMGR\RELEASED\04691CC\source\04693e.ddbDrawn By:VIN5GND3SW4

Pagina 283

12345 6ABCD654321DCBATitleNumber RevisionSizeBDate: 13-Jan-2005 Sheet of File: N:\PCBMGR\UNREL\04930PW\Protel\04930.DDBDrawn By:12J2+15TP2TP3C160.1

Pagina 284 - AND MAINTENANCE

11223344D DC CB BA ATi tleNumber Revis ionSizeADate: 6/28/2004 Sheet of File: N:\PCBMGR\..\04468B.sch Drawn By:12345678JP1No t U sedR122R2SCH, E-S

Pagina 285

11223344D DC CB BA ATitleNumber RevisionSizeADate: 1/28/2010 Sheet ofFil e: N:\PCBMGR\..\06731-1_ETHERNET.SchDocDrawn By:Auxiliary I/O Board

Pagina 286 - DO NOT use pink-poly bags

11223344D DC CB BA ATitleNum ber RevisionSizeADate: 1/28/2010 Sh eet ofFil e: N:\PCBMGR\..\06731-2_USB.SchDoc Drawn By:Auxiliary I/O Board

Pagina 287 - GLOSSARY

11223344D DC CB BA ATitleNumber RevisionSizeADate: 1/28/2010 Sheet ofFi le: N:\PCBM G R \..\06731-3_ ADC.Sch Doc Drawn By:Auxiliary I/O Boar

Pagina 288

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\..\06696.P1.R3.schdoc Drawn By:GUI Interface06698D14

Pagina 289

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\ ..\06696.P2.R3.schdoc D rawn By:06698D2422uHL1Vin5SH

Pagina 290

Getting Started Teledyne API – T101 Operation Manual 34 Table 3-3. Analog Input Pin Assignments PIN DESCRIPTION DAS PARAMETER1 1 Analog input # 1

Pagina 291

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\..\06696.P3.R3.schdoc Drawn By:06698D4312345To old

Pagina 292

112233445566D DC CB BA ATitleNumber RevisionSizeBDate: 6/24/2010 Sheet ofFile: N:\PCBMGR\ ..\06696.P4.R3.schdoc D rawn By:06698D44TOUCH SCREE

Pagina 293

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Pagina 294 - (Primary Setup Menu)

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Pagina 295

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Pagina 296

   > $$>+?@ !FGJMQ XZX[ &?Q Q\  !\ !8 X X G

Pagina 297 - ENTER SETUP PASS: 818

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Pagina 298

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Pagina 299 - REPORTED

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Pagina 300

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Pagina 301

Teledyne API – T101 Operation Manual Getting Started 35 Table 3-4. Analog Output Pin Assignmentss PIN ANALOG OUTPUT VOLTAGE OUTPUT CURRENT LOOP

Pagina 302

    $%&''&%$7LWOH1XPEHU 5HYLVLRQ6L]H2UFDG%'DWH 0D\ 6KHHWRI)LOH 1?3&%0*5?5-(VHULHV0RWKH

Pagina 303

11223344D DC CB BA ATitleNumber RevisionSizeADate: 5/7/2010 Sheet ofFil e: N:\PCBMGR\..\06882-P1-R0.SchDoc Drawn By:LVDS, Transmitter BoardB

Pagina 304

Getting Started Teledyne API – T101 Operation Manual 36 3.3.3. CONNECTING THE STATUS OUTPUTS The analyzer’s status outputs are accessed through a

Pagina 305

Teledyne API – T101 Operation Manual Getting Started 37 Table 3-5. Status Output Signals REAR PANEL LABEL STATUS DEFINITION CONDITION 1 SYSTEM OK

Pagina 306

ii This page intentionally left blank. 07266B DCN6485

Pagina 307

Getting Started Teledyne API – T101 Operation Manual 38 SPAN CALZERO CALSPAN CALZERO CALCONTROL IN A B C D E F U

Pagina 308

Teledyne API – T101 Operation Manual Getting Started 39 3.3.5. CONNECTING THE COMMUNICATIONS PORTS 3.3.5.1. Connecting the Serial Ports To utilize

Pagina 309

Getting Started Teledyne API – T101 Operation Manual 40 Table 3-7. Inlet / Outlet Connector Descriptions REAR PANEL LABEL FUNCTION SAMPLE Co

Pagina 310

Teledyne API – T101 Operation Manual Getting Started 41 ValveNeedle valve to control flow Chassis SampleZero Air Ex haustSpanMODEL 701 Zero Air Ge

Pagina 311

Getting Started Teledyne API – T101 Operation Manual 42 PMT EXHA UST THROUGH OUTERLAYER OF KICK ER REACTION CELL PURGE FL O W SENSORFLOW /

Pagina 312

Teledyne API – T101 Operation Manual Getting Started 43 CAUTION The exhaust from the external pump needs to be vented outside the immediate area o

Pagina 313

Getting Started Teledyne API – T101 Operation Manual 44 Table 3-9. NIST-SRM's Available for Traceability of H2S & SO2 Calibration Gases N

Pagina 314

Teledyne API – T101 Operation Manual Getting Started 45 3.4.1.1. Connections with Internal Valve Options Installed This section covers pneumatic co

Pagina 315

Getting Started Teledyne API – T101 Operation Manual 46 ZERO/SPAN (Z/S) VALVE GAS FLOW PMT EXHAUST TO OUTER LAYER OF KICKER REACTION CELL PURG

Pagina 316

Teledyne API – T101 Operation Manual Getting Started 47 The state of the zero/span valves can also be controlled:  Manually from the analyzer’s f

Pagina 317

iii SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage.

Pagina 318

Getting Started Teledyne API – T101 Operation Manual 48 Temperature of the H2S: Increasing the temperature of the increases the pressure inside th

Pagina 319

Teledyne API – T101 Operation Manual Getting Started 49 PMT EXHAUST TO OUTER LAYER OF KIC KER REA CTION CELL PUR GE FLOW SENSORFLOW / PRESSURE

Pagina 320 - Name Description

Getting Started Teledyne API – T101 Operation Manual 50 The state of the IZS valves can also be controlled:  Manually from the analyzer’s front

Pagina 321

Teledyne API – T101 Operation Manual Getting Started 51 Table 3-12. Possible Warning Messages at Start-Up WARNING MESSAGE MEANING ANALOG CAL WARNI

Pagina 322 - Name Description Units

Getting Started Teledyne API – T101 Operation Manual 52 3.5.4. FUNCTIONAL CHECK 1. After the analyzer’s components have warmed up for at least 30

Pagina 323

Teledyne API – T101 Operation Manual Getting Started 53 3.6. INITIAL CALIBRATION 3.6.1. BASIC CALIBRATION PROCEDURE The following three-step proced

Pagina 324

Getting Started Teledyne API – T101 Operation Manual 54 STEP TWO: Set the expected H2S span gas concentration. This sequence causes the analyzer

Pagina 325

Teledyne API – T101 Operation Manual Getting Started 55 STEP THREE: Perform the zero/span calibration procedure: M-P CAL H2S STB=X

Pagina 326 - (single-bit; read/write)

Getting Started Teledyne API – T101 Operation Manual 56 3.6.2. INTERFERENCES FOR H2S MEASUREMENTS It should be noted that the fluorescence method

Pagina 327

57 4. OPERATING INSTRUCTIONS The T101 is a micro-computer-controlled analyzer with a dynamic menu interface for easy and yet powerful and flexible

Pagina 328

iv CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures cor

Pagina 329 - T101 Spare Parts List

Operating Instructions Teledyne API – T101 Operation Manual 58 MODE DESCRIPTION SETUP X.#2 SETUP mode is being used to configure the analyzer. Th

Pagina 330

Teledyne API – T101 Operation Manual Operating Instructions 59 Table 4-2. Test Functions Defined DISPLAY PARAMETER UNITS DESCRIPTION RANGE Range

Pagina 331

Operating Instructions Teledyne API – T101 Operation Manual 60 TEMP Converter Temperature H2S present in the sample gas into SO2. TEST2 Test Sign

Pagina 332 - 07266B DCN6845

Teledyne API – T101 Operation Manual Operating Instructions 61 4.2.2. WARNING MESSAGES The most common instrument failures will be reported as a war

Pagina 333 - T101, M101E

Operating Instructions Teledyne API – T101 Operation Manual 62 To view and clear warning messages: SAMPLE HVPS WARNING

Pagina 334

Teledyne API – T101 Operation Manual Operating Instructions 63 The default status of the calibration password is OFF. To enable the calibration pas

Pagina 335

Operating Instructions Teledyne API – T101 Operation Manual 64 4.4. SETUP MODE The SETUP mode allows you to configure the analyzer’s hardware and

Pagina 336

Teledyne API – T101 Operation Manual Operating Instructions 65 NOTE If the analyzer beeps when you press the EXIT button, it means that you’ve made

Pagina 337

Operating Instructions Teledyne API – T101 Operation Manual 66 4.4.2. SETUP – ACAL: AUTO CALIBRATION Used to set up and operate the internal valve

Pagina 338

Teledyne API – T101 Operation Manual Operating Instructions 67 AUTO mode, see section 4.4.4.6) or operate independently (IND mode, see Section 4.4.

Pagina 339

v WARRANTY WARRANTY POLICY (02024 F) Teledyne Advanced Pollution Instrumentation (TAPI), a business unit of Teledyne Instruments, Inc., provides th

Pagina 340

Operating Instructions Teledyne API – T101 Operation Manual 68 4.4.4.3. Reporting Range Modes The T101 provides three analog output range modes to

Pagina 341

Teledyne API – T101 Operation Manual Operating Instructions 69 4.4.4.4. Single Range Mode (SNGL) The default range mode for the analyzer is single r

Pagina 342

Operating Instructions Teledyne API – T101 Operation Manual 70 4.4.4.5. Independent Range Mode (IND) Selecting independent range mode allows the A

Pagina 343

Teledyne API – T101 Operation Manual Operating Instructions 71 NOTE On analyzers with the multigas option activated (see Sections 4.5.1 and 6.8) th

Pagina 344

Operating Instructions Teledyne API – T101 Operation Manual 72 NOTE On analyzers with the multigas option activated (see Section s 4.5.1 and 6.8)

Pagina 345

Teledyne API – T101 Operation Manual Operating Instructions 73 4.4.4.8. Dilution Ratio The dilution ratio is a software option that allows the user

Pagina 346

Operating Instructions Teledyne API – T101 Operation Manual 74 4.4.5. SETUP – PASS: PASSWORD PROTECTION The menu system provides password protecti

Pagina 347

Teledyne API – T101 Operation Manual Operating Instructions 75 4.4.6. SETUP – CLK: SETTING THE INTERNAL TIME-OF-DAY CLOCK The T101 has a built-in cl

Pagina 348

Operating Instructions Teledyne API – T101 Operation Manual 76 In order to compensate for CPU clocks which run fast or slow, there is a variable t

Pagina 349

Teledyne API – T101 Operation Manual Operating Instructions 77 4.5. SETUP – VARS: USING THE INTERNAL VARIABLES The T101 has several-user adjustable

Pagina 350

vi This page intentionally left blank. 07266B DCN6485

Pagina 351

Operating Instructions Teledyne API – T101 Operation Manual 78 NO. VARIABLE DESCRIPTION ALLOWED VALUES 11 TIME_SINCE_SVC Displays time in hours s

Pagina 352

Teledyne API – T101 Operation Manual Operating Instructions 79 To access and navigate the VARS menu, use the following touchscreen button sequence:

Pagina 353

Operating Instructions Teledyne API – T101 Operation Manual 80 4.5.1. SETTING THE GAS MEASUREMENT MODE If the switching valves software is activat

Pagina 354

Teledyne API – T101 Operation Manual Operating Instructions 81 4.6. SETUP – DIAG: USING THE DIAGNOSTICS FUNCTIONS A series of diagnostic tools is g

Pagina 355

Operating Instructions Teledyne API – T101 Operation Manual 82 To access the DIAG functions press the following buttons: SAMPLE ENTE

Pagina 356 - Auxiliary I/O Board (USB)

Teledyne API – T101 Operation Manual Operating Instructions 83 4.6.1. SIGNAL I/O The signal I/O diagnostic mode allows reviewing and changing the di

Pagina 357 - Auxiliary I/O Board (ADC)

Operating Instructions Teledyne API – T101 Operation Manual 84 4.6.2. ANALOG OUTPUT STEP TEST This test can be used to check the accuracy and prop

Pagina 358 - GUI Interface

Teledyne API – T101 Operation Manual Operating Instructions 85 4.6.3. ANALOG I/O CONFIGURATION Table 4-8 lists the analog I/O functions that are ava

Pagina 359

Operating Instructions Teledyne API – T101 Operation Manual 86 The following DC current output limits apply to the current loop modules: Table 4-1

Pagina 360 - MBUS B B

Teledyne API – T101 Operation Manual Operating Instructions 87 4.6.3.1. Analog Output Signal Type and Range Span Selection To select an output signa

Pagina 361

vii ABOUT THIS MANUAL This T101 operation manual, PN 07266, is comprised of multiple documents in PDF format, as listed below. Part No. Rev Name/

Pagina 362

Operating Instructions Teledyne API – T101 Operation Manual 88 4.6.3.2. Analog Output Calibration Mode Analog output calibration should be carried

Pagina 363

Teledyne API – T101 Operation Manual Operating Instructions 89 To automatically calibrate an single analog channel, activate the ANALOG I/O CONFIGUR

Pagina 364

Operating Instructions Teledyne API – T101 Operation Manual 90 To select manual output calibration for a particular channel, activate the ANALOG I

Pagina 365

Teledyne API – T101 Operation Manual Operating Instructions 91 V +DC Gnd V OUT + V OUT -V IN + V IN - Recording Device ANALYZERSee Table 3-1

Pagina 366

Operating Instructions Teledyne API – T101 Operation Manual 92 4.6.3.4. Analog Output Offset Adjustment Some analog signal recorders require that

Pagina 367

Teledyne API – T101 Operation Manual Operating Instructions 93 Similar to the voltage calibration, the software allows this current adjustment to be

Pagina 368

Operating Instructions Teledyne API – T101 Operation Manual 94 To adjust the zero and span values of the current outputs, activate the ANALOG I/O

Pagina 369

Teledyne API – T101 Operation Manual Operating Instructions 95 4.6.3.6. AIN Calibration This is the sub-menu to conduct the analog input calibration

Pagina 370

Operating Instructions Teledyne API – T101 Operation Manual 96 4.6.3.7. Analog Inputs (XIN1…XIN8) Option Configuration To configure the analyzer’s

Pagina 371 - To LCD Display

Teledyne API – T101 Operation Manual Operating Instructions 97 4.6.4. OPTIC TEST The optic test function tests the response of the PMT sensor by tur

Comentarios a estos manuales

Sin comentarios