Teledyne API – T101 Operation Manual Operating Instructions
141
4.8.2.11. HOLDOFF Feature
The DAS HOLDOFF feature allows to prevent data collection during calibrations and
during the DAS_HOLDOFF period enabled and specified in the VARS (Section 4.8). To
enable or disable the HOLDOFF, follow the instruction shown in Section 4.11.2.2 then
press:
Edit Data Ch annel Menu
From th e DATA ACQUISITION menu
SETUP X.X 0) CONC: ATIMER, 2, 4032, R
PREV NEXT INS DEL EDIT PRNT EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X NAME:CONC
<SET SET> EDIT PRINT EXIT
SETUP X.X CAL HOLD OFF:ON
SET> EDIT PRINT EXIT
ENTR accepts the new
setting and retu rns to the
pre vi ous menu.
EXIT ignor es the n ew setting
and returns to the previo us
menu.
SETUP X.X CAL HOLD OFF:ON
ON ENTR EXIT
Toggle to turn HOLDOFF
ON or OFF
Press SET> until…
07266B DCN6485