Troubleshooting & Service Model T101 Instruction Manual
218
Toggles parameter
ON/OFF
Exit returns to
DIAG display & all values
return to software control
DIAG SIGNAL I/O
PREV NEXT ENTR EXIT
DIAG I/O 0 ) EXT_ZERO_CAL=ON
PREV NEXT JUMP PRNT EXIT
DIAG I/O 23) ST H2S_MODE=ON
PREV NEXT JUMP ON PRNT EXIT
D IAG I/O 3 7) SAMPLE_PR ESSURE=6000.0 MV
PREV NEXT JUMP PRNT EXIT
DIAG I/O 23) ST H2S_MODE OFF
PREV NEXT JUMP OFF PRNT EXIT
If parameter is an
in put sig nal
If parameter is an output
signal or control
SETUP X .X SECONDARY SETUP MENU
COMM VARS DIAG EX IT
SAMPLE ENTER SETUP PASS : 818
8 1 8 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE = 500.0 PPB H2S =XXX.X
< TST TST > CAL SETUP
Figure 9-2. Example of Signal I/O Function
9.1.4. STATUS LEDS
Several color-coded, light-emitting diodes (LEDs) are located inside the instrument to
determine if the analyzer’s CPU, I
2
C communications bus and relay board are
functioning properly.