Teledyne T802 Manual de usuario Pagina 173

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 314
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 172
Teledyne API T802 Paramagnetic O
2
Analyzer Operation Manual Calibration Procedures
07275A (DCN6005) 171
Cylinders of calibrated O
2
gas traceable to NIST-Standard Reference Material
specifications (also referred to as SRMs or EPA protocol calibration gases) are
commercially available.
Table 9-1: NISTSRM's Available for Traceability of O
2
Calibration Gases
NIST-SRM Type Nominal Concentration
2657a
O
2
in N
2
2%
2658a
O
2
in N
2
10 %
2659a
O
2
in N
2
21%
2619a
1
CO
2
in N
2
0.5%
2620a
1
CO
2
in N
2
1%
2622a
1
CO
2
in N
2
2%
2624a
1
CO
2
in N
2
3%
2744b
1
CO
2
in N
2
7%
2745
1
CO
2
in N
2
16%
1
Used to calibrate optional CO
2
sensor.
Note
For span point calibration it is generally a good idea to use 80% of the reporting
range for that channel. For instance, if the reporting range of the instrument is
set for 5%, the proper span gas would be 4%
.
9.1.3. DATA RECORDING DEVICES
A strip chart recorder, data acquisition system or digital data acquisition system
should be used to record data from the serial or analog outputs of the T802.
07275B DCN6418
Vista de pagina 172
1 2 ... 168 169 170 171 172 173 174 175 176 177 178 ... 313 314

Comentarios a estos manuales

Sin comentarios