Teledyne 200A Manual de usuario

Busca en linea o descarga Manual de usuario para De medición, control y control Teledyne 200A. Teledyne 200A User's Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 225
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
TELEDYNE
INSTRUMENTS
Advanced Pollution Instrumentation
A Teledyne Technologies Company
INSTRUCTION MANUAL
Chemiluminescent
NO/NO
2
/NO
X
ANALYZER
Model 200A
© Teledyne Advanced Pollution Instrumentation
(T-API)
9480 Carroll Park Drive
San Diego, CA 92121-5201
TOLL-FREE: 800-324-5190
FAX: 858-657-9816
TEL: 858-657-9800
WEB SITE: www.teledyne-api.com
Vista de pagina 0
1 2 3 4 5 6 ... 224 225

Indice de contenidos

Pagina 1 - TELEDYNE

TELEDYNE INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies Company INSTRUCTION MANUAL Chemiluminescent N

Pagina 2 - HIGHLIGHTS

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.6 Moly Converter Efficiency ...

Pagina 3

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.3 Manual Zero/Span Check with IZS Option The Internal Zero/Span (

Pagina 4 - NITROGEN OXIDE ANALYZER

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-8: Enabling Dynamic Zero/Span StepNumber Action Comment

Pagina 5

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.6 Calibrate on NO2 Permeation Tube If the M200A is equipped with

Pagina 6 - SAFETY MESSAGES

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-9: Enabling CAL-ON-NO2 Step Number Action Comment 1. Pr

Pagina 7

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.7 Use of Zero/Span Valves or IZS with Remote Contact Closure The

Pagina 8 - TABLE OF CONTENTS

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8 EPA Protocol Calibration If the M200A is to be used for EPA com

Pagina 9

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-11: Activity Matrix for Calibration Equipment and Supplie

Pagina 10

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-12: Activity Matrix for Calibration Procedure Equipment/

Pagina 11

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.2.2 Selection of NO Span Gas Standards The NIST-SRM's prov

Pagina 12 - LIST OF FIGURES

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.4 Gas Phase Titration (GPT) System 7.8.4.1 GPT Principle of Op

Pagina 13 - LIST OF TABLES

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.2.11 Slow Response...

Pagina 14

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 1. Select an NO standard gas that has a nominal concentration in th

Pagina 15

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 NOTE If tr is >2 minutes or if PR is <2.75 ppm-min, changes i

Pagina 16 - 1 HOW TO USE THIS MANUAL

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Because low flows are difficult to control and measure, it is often

Pagina 17

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.5 Dynamic Multipoint Calibration Procedure The procedure for ca

Pagina 18 - 2 GETTING STARTED

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-13: Diagram of GPT Calibration System 7-21 PRINTED DOCUM

Pagina 19

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.5.1 Zero Calibration Procedure Since the zero gas concentration

Pagina 20

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-15: Expected Span Gas Concentration Procedure Step Number

Pagina 21

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Plot the analyzer NO and NOx responses versus the corresponding calc

Pagina 22

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 4. Calculate the resulting NO2 concentrations as follows: F]NO[*F

Pagina 23 - 2.3 Initial Operation

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.6.1 Automatic Moly Converter Efficiency Compensation The M200A

Pagina 24

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 LIST OF FIGURES FIGURE 2-1: REMOVAL OF SHIPPING SCREWS & CHECK

Pagina 25

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.7 Calibration Frequency To ensure accurate measurements of the N

Pagina 26

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-18: Definition of Level 1 and Level 2 Zero and Span Check

Pagina 27

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.9 Summary of Quality Assurance Checks The following items shoul

Pagina 28

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.10 ZERO and SPAN Checks A system of Level 1 and Level 2 zero/sp

Pagina 29

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.10.2 Precision Check A periodic check is used to assess the dat

Pagina 30

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-20: NIST-SRM's Available for Traceability of Calibra

Pagina 31

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.8.12.1 NO Working Standard Traced to NIST NO Standard First, use

Pagina 32 - WARRANTY

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.9 Calibration of Independent Ranges or Autoranging There are addi

Pagina 33

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.10 Calibration Quality After calibration is complete, it is very

Pagina 34 - 3.3 Warranty

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-21: Calibration Quality Check Step Number Action Commen

Pagina 35 - TERMS AND CONDITIONS

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 LIST OF TABLES TABLE 2-2-1: REMOVAL OF SHIPPING SCREWS & CHECK

Pagina 36 - ANALYZER

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.11 References 1. Environmental Protection Agency, Title 40, Code

Pagina 37

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7-38 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTE

Pagina 38 - 4.2 Operation Summary

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 8 MAINTENANCE 8.1 Maintenance Schedule NOTE The operations outline

Pagina 39 - 4.2.5 Front Panel

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 8-2 PRINTED DOCUMENTS ARE UNCONTROLLED Table 8-8-2: Preventative Ma

Pagina 40

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 8.2 Replacing the Sample Particulate Filter The particulate filter

Pagina 41 - 4.2.6 Power Supply Module

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 8-8-3: Replacing the Particulate Filter 8-4 PRINTED DOCUMENT

Pagina 42

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 8.3 Replacing the Ozone Scrubber Charcoal The Ozone scrubber consis

Pagina 43

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 NOTE If the instrument is to be turned off for more than 8 hours, t

Pagina 44 - 4.2.9 Molydbenum Converter

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 8.6 Replacing the Molybdenum Converter The Molybdenum Converter is

Pagina 45

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 8-8-4: Molybdenum Converter Assembly 8-8 PRINTED DOCUMENTS

Pagina 46 - 5 SOFTWARE FEATURES

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 TABLE 7-7-4: ENTER EXPECTED SPAN GAS CONCENTRATIONS PROCEDURE ...

Pagina 47

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 8.7 Cleaning the Reaction Cell The reaction cell should be cleaned

Pagina 48

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 OZONE INLET FITTINGSAMPLE INLET FITTINGSPRINGFILTERORINGORIFICE 10 M

Pagina 49 - 5.1.1 Sample Menu

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 8.8 Pneumatic Line Inspection Particulate matter in the pneumatic l

Pagina 50 - 5.1.2 Set-Up Menu

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 8-8-6: Pneumatic Diagram - Standard Configuration 8-12 PRIN

Pagina 51

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 8-8-7: Pneumatic Diagram with Zero/Span Valve Option 8-13 PR

Pagina 52

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 8-8-8: Pneumatic Diagram with IZS Option 8-14 PRINTED DOCUME

Pagina 53 - 5.2 Sample Mode

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 8.9 Leak Check Procedure If a leak checker is not available, it is

Pagina 54

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 8-16 PRINTED DOCUMENTS ARE UNCONTROLLED 8.10 Light Leak Check Proce

Pagina 55

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9 TROUBLESHOOTING, ADJUSTMENTS NOTE The operations outlined in thi

Pagina 56

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 3. Check the TEST functions: A. Compare the TEST functions to the

Pagina 57

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 x PRINTED DOCUMENTS ARE UNCONTROLLED TABLE 9-9-12: V/F BOARD SETTIN

Pagina 58

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.1 Operation Verification-M200A Diagnostic Techniques 9.1.1 Fault

Pagina 59 - 5.3 Set-Up Mode

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-1: Test Functions Test Function Factory Set-Up Comment

Pagina 60

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up

Pagina 61 - 5.3.4 Range Menu

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up

Pagina 62

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up

Pagina 63

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.1.2 Fault Diagnosis with WARNING Messages The M200A monitors seve

Pagina 64

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-2: Front Panel Warning Messages Message Description SYST

Pagina 65 - 5.3.8 Communications Menu

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-2: Front Panel Warning Messages (Continued) Message Descri

Pagina 66 - 5.3.9 Variables Menu (VARS)

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.1.3 Fault Diagnosis Using DIAGNOSTIC Mode Diagnostic mode can be

Pagina 67 - 5.4 Status Output

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-3: Summary of Diagnostic Modes DIAG Mode Description SIG

Pagina 68 - 5.5 RS-232 Interface

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 1 HOW TO USE THIS MANUAL The Model 200A has been designed to provid

Pagina 69

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.1.3.1 Signal I/O Diagnostic Table 9-9-4: Diagnostic Mode - Signa

Pagina 70

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Co

Pagina 71

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Co

Pagina 72

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Co

Pagina 73 - 5.5.2 Command Summary

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 3. The value in PMT should come up to 2000 mV ± 500 mV in less than

Pagina 74

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.1.3.5 Analog Out Step Test The Step Test is used to test the func

Pagina 75

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 The reaction cell pressure compensation factor is equal to 1.0000 wh

Pagina 76

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-5: Model 200A Internal Variables No. Name Units Default

Pagina 77

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-6: Test Channel Readings TEST Channel Minimum* Maximum

Pagina 78

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-6: Test Channel Readings (Continued) TEST Channel Minimum*

Pagina 79

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 1-2 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTEN

Pagina 80

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.1.6 Factory Calibration Procedure The Factory Cal procedure balan

Pagina 81

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9. Calculate the expected NORM PMT mV reading. For ranges up to 20

Pagina 82

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-7: Span Calibration Voltage 9-25 PRINTED DOCUMENTS ARE

Pagina 83

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Electric Test (ET) Procedure: 1. To adjust ET press SETUP-MORE-DIAG

Pagina 84 - 5.5.8 Internal Variables

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 4. Change in span gas concentration. A. Zero air or ambient air le

Pagina 85

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.2.3 No Response to Sample 1. Confirm correct operation of analo

Pagina 86 - 6.2 Zero/Span Valves

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.2.4 Negative Output 1. Mis-calibration. The 'zero' gas

Pagina 87

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.2.6 Unstable Span Common causes are: 1. Leak in pneumatic system

Pagina 88 - Mode No. Mode Name Action

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.2.7 Unstable Zero Common causes are: 1. Leak in pneumatic system

Pagina 89

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.2.9 Inability to Zero If the ZERO button is not illuminated when

Pagina 90 - 6.5 Permeation Tube

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 2 GETTING STARTED 2.1 Unpacking 1. Verify that there is no appare

Pagina 91

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.2.10 Non-Linear Response Common causes are: 1. Leak in pneumatic

Pagina 92

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.2.11 Slow Response 1. Contaminated or dirty sample delivery pneu

Pagina 93

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3 Subsystem Troubleshooting and Adjustments 9.3.1 Computer, Disp

Pagina 94

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-8: CPU Board Jumper Settings 9-36 PRINTED DOCUMENTS ARE

Pagina 95

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Like the display, the overall functioning of the CPU can be confirme

Pagina 96

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.2 RS-232 Communications The M200A uses the RS-232 communication

Pagina 97

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 To troubleshoot a modem connection first disconnect the RS-232 cable

Pagina 98

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-9: RS-232 PIN Assignments 9-40 PRINTED DOCUMENTS ARE UN

Pagina 99 - Zero/Span Valves Option

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.3 Voltage/Frequency (V/F) Board The V/F Board consists of 16 an

Pagina 100

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 4. Use a DVM on each of the analog output channels to confirm the c

Pagina 101 - NO and NO

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 WARNING Analyzer Exhaust – O3 Scrubber – Pump Pack Danger – Analyz

Pagina 102 - Permeation Tube

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.3.1 ADC/DAC Calibration Procedure Due to the stability of moder

Pagina 103

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-11: V/F Board Switch Settings DAC # Sw 1 Sw 2 Sw 3 Sw

Pagina 104 - Closure

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 1. After the switches and jumpers are set, turn on instrument power

Pagina 105

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-12: V/F Board Settings 9-46 PRINTED DOCUMENTS ARE UNCONT

Pagina 106

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.3.2 Changing Output Voltage Ranges Several different output vol

Pagina 107

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 In addition there is a thermocouple amplifier for the Moly converter

Pagina 108 - 7.8.3 Data Recording Device

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.4.3 Status Output Lines, External Contact Closures The Status l

Pagina 109 - 223 ONOONO 

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-13: Power Supply Module Subassemblies Module Description

Pagina 110 - = analyzer flow demand (cm

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-14: Power Supply Module Layout 9-51 PRINTED DOCUMENTS AR

Pagina 111 - = 500 cm

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-15: Electrical Block Diagram 9-52 PRINTED DOCUMENTS ARE

Pagina 112

Teledyne API, Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 HIGHLIGHTS The purpose of this Highlights Page is to list the

Pagina 113

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 2-2-1: Removal of Shipping Screws & Check for Correct Po

Pagina 114

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 PSM Diagnostic Procedures The Linear Power Supply board can be teste

Pagina 115

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.6 Ozone Generator CAUTION - DANGER Lethal voltages present insi

Pagina 116

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-17: Ozone Generator Control Conditions Condition Descrip

Pagina 117

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-18: Ozone Generator Subsystem 9-56 PRINTED DOCUMENTS ARE

Pagina 118

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.7 Flow/Pressure Sensor The flow/pressure sensor board consists

Pagina 119

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 1. In SAMPLE mode, scroll the TEST functions to SAMP FLW. Observe t

Pagina 120 - 7.8.7 Calibration Frequency

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-19: Flow/Pressure Sensor 9-59 PRINTED DOCUMENTS ARE UNCO

Pagina 121

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-20: NOx Sensor Module 9-60 PRINTED DOCUMENTS ARE UNCONTR

Pagina 122

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-21: NOx Sensor Module 9-61 PRINTED DOCUMENTS ARE UNCONTR

Pagina 123 - 7.8.10 ZERO and SPAN Checks

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.8 NOx Sensor Module Table 9-9-20 and Table 9-9-21 show the asse

Pagina 124

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 2-2-2: Rear Panel 2-4 PRINTED DOCUMENTS ARE UNCONTROLLED

Pagina 125

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-22: PMT Cooler Subsystem 9-63 PRINTED DOCUMENTS ARE UNCO

Pagina 126

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.8.4 PMT Cooler The PMT cooler uses a Peltier cooler supplied wi

Pagina 127

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 9-9-23: High Voltage Power Supply 9-65 PRINTED DOCUMENTS ARE

Pagina 128 - 7.10 Calibration Quality

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.9 Z/S Valves & IZS Permeation Tube Oven The Z/S Valves and

Pagina 129

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9.3.10 Pneumatic System The pneumatic system is diagramed in Table

Pagina 130 - 7.11 References

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 9-68 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTE

Pagina 131

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 10-1 PRINTED DOCUMENTS ARE UNCONTROLLED 10 M200A SPARE PARTS LIST T

Pagina 132 - 8 MAINTENANCE

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 10-2 PRINTED DOCUMENTS ARE UNCONTROLLED Part No. Description 011340

Pagina 133

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 10-3 PRINTED DOCUMENTS ARE UNCONTROLLED Part No. Description 052930

Pagina 134

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 10-4 PRINTED DOCUMENTS ARE UNCONTROLLED Part No. Description OR0000

Pagina 135

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 2-2-3: Inlet and Exhaust Venting 2-5 PRINTED DOCUMENTS ARE U

Pagina 136

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 10-5 PRINTED DOCUMENTS ARE UNCONTROLLED Table 10-2: Teledyne API MO

Pagina 137

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 10-6 PRINTED DOCUMENTS ARE UNCONTROLLED Table 10-4: Teledyne API MO

Pagina 138

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 10-7 PRINTED DOCUMENTS ARE UNCONTROLLED Table 10-6: Teledyne API MO

Pagina 139

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 10-8 PRINTED DOCUMENTS ARE UNCONTROLLED Part No. Description 014020

Pagina 140

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 APPENDIX A ELECTRICAL SCHEMATICS Table A-1: Electrical Schematics

Pagina 141

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 A-2 PRINTED DOCUMENTS ARE UNCONTROLLED Table A-1: Electrical Schema

Pagina 142 - Teledyne API Model 200A NO

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 2.3 Initial Operation 1. Turn on the instrument power. 2. The di

Pagina 143

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7. When the instrument is warmed up, re-check the TEST functions ag

Pagina 144

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Step 2 - Calibrate the instrument: Zero/Span Calibration Procedure S

Pagina 145

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Step 3 - Review the quality of the calibration: Calibration Quality

Pagina 146 - 8.9 Leak Check Procedure

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 2-2-4: Front Panel 2-10 PRINTED DOCUMENTS ARE UNCONTROLLED

Pagina 147

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 2-2-5: Assembly Layout 2-11 PRINTED DOCUMENTS ARE UNCONTROLL

Pagina 148

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 2-12 PRINTED DOCUMENTS ARE UNCONTROLLED Table 2-2-6: Final Test and

Pagina 149

THIS PAGE IS INTENTIONALLY LEFT BLANK

Pagina 150

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 2-1: Final Test and Calibration Values (Continued) Parameter

Pagina 151 - (table continued)

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 2-14 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTE

Pagina 152

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 3 SPECIFICATIONS, AGENCY APPROVALS, WARRANTY 3.1 Specifications Ra

Pagina 153

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 3.2 EPA Equivalency Designation Advanced Pollution Instrumentation,

Pagina 154

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 3.3 Warranty ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) (

Pagina 155

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 3-4 PRINTED DOCUMENTS ARE UNCONTROLLED TERMS AND CONDITIONS All un

Pagina 156

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 4 THE M200A NOX ANALYZER 4.1 Principle of Operation The Teledyne A

Pagina 157

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 4-4-1: Block Diagram 4-2 PRINTED DOCUMENTS ARE UNCONTROLLED

Pagina 158

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 4.2 Operation Summary 4.2.1 Sensor Module, Reaction Cell, Detector

Pagina 159

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 4.2.3 Computer Hardware and Software CPU Board The M200A Analyzer i

Pagina 160

TELEDYNE INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies Company INSTRUCTION MANUAL MODEL 200A NITROGEN OXIDE ANA

Pagina 161

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 The center field displays TEST values. The TEST functions allow you

Pagina 162

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 4.2.6 Power Supply Module The Power supply module supplies AC and D

Pagina 163

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 NOTE On vacuum vs absolute pressure: Many vacuum gauges read relati

Pagina 164

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 4-4-3: External Pump Pack 4-8 PRINTED DOCUMENTS ARE UNCONTRO

Pagina 165

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 4.2.8 Ozone Generator Because of the instability of ozone, it is ne

Pagina 166

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 4-10 PRINTED DOCUMENTS ARE UNCONTROLLED THIS PAGE IS INTE

Pagina 167

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5 SOFTWARE FEATURES The M200A control software has two major operat

Pagina 168

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-1: Sample Menu Tree 5-2 PRINTED DOCUMENTS ARE UNCONTROL

Pagina 169

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-2: Setup Menu Tree 5-3 PRINTED DOCUMENTS ARE UNCONTROLLE

Pagina 170

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.1.1 Sample Menu Table 5-5-3: M200A Sample Menu Structure Menu Le

Pagina 171

THIS PAGE IS INTENTIONALLY LEFT BLANK

Pagina 172

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.1.2 Set-Up Menu Table 5-5-4: M200A Setup Menu Structure Setup Me

Pagina 173 - 9.2 Performance Problems

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-5: M200A Menu Structure - Setup Menu #2 Setup Menu #2 Lev

Pagina 174 - 9.2.2 Flow Check

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Setup Menu #3 Level 1 Level 2 Level 3 Level 4 Description Refere

Pagina 175

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.2 Sample Mode The M200A software has two major operating modes, t

Pagina 176 - 9.2.5 Excessive Noise

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Stability The instrument noise is computed using the standard deviat

Pagina 177 - 9.2.6 Unstable Span

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 High Voltage Power Supply (HVPS) The HVPS reading is a measure of th

Pagina 178 - 9.2.8 Inability to Span

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Moly Temperature The moly temperature is controlled by the computer.

Pagina 179 - 9.2.9 Inability to Zero

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.2.2 CAL, CALS, CALZ, Calibration Functions The calibration and ze

Pagina 180 - 9.2.10 Non-Linear Response

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.2.2.5 Formula Values The slope and offset terms should be checked

Pagina 181 - 9.2.11 Slow Response

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.2.2.6 Automatic Converter Efficiency Compensation The M200A can a

Pagina 182

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 SAFETY MESSAGES Your safety and the safety of others is very importa

Pagina 183

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 The M200A comes pre-programmed with a set of useful Data Channels fo

Pagina 184

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-7: DAS Data Channel Editing Step Action Comment 1. Pres

Pagina 185 - 9.3.2 RS-232 Communications

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.3.4.1 Single Range This range option selects a single range for a

Pagina 186

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 To use the Dilution feature: 1. SELECT UNITS For stack measurement

Pagina 187

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.3.4.5 Concentration Units The M200A can display concentrations in

Pagina 188

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.3.5 Password Enable If password protection is enabled, a password

Pagina 189

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.3.9 Variables Menu (VARS) This menu enables you to change the set

Pagina 190

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-9: M200A Operating Modes Mode Description ZERO CAL D Au

Pagina 191 - *Required settings

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-10: Status Output Pin Assignments Output # Pin # Defini

Pagina 192

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.5.1 Setting Up the RS-232 Interface The baud rate is set from the

Pagina 193

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 THIS PAGE IS INTENTIONALLY LEFT BLANK ii PRINTED DOCUMEN

Pagina 194 - 9.3.4 Status/Temp Board

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-11: RS-232 Port Setup - Front Panel Decimal Value Descri

Pagina 195

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-12: RS-232 Switching From Terminal Mode to Computer Mode

Pagina 196 - 9.3.5 Power Supply Module

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Security Feature The RS-232 port is often connected to a public tele

Pagina 197

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Communication Protocols The RS-232 port can communicate in the follo

Pagina 198

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-14: RS-232 Command Summary Commands Definition ? [id] P

Pagina 199

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-15: RS-232 Command Summary Terminal Mode Editing Keys De

Pagina 200 -  C. The

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-16: RS-232 Interface Command Types First Character Messa

Pagina 201 - 9.3.6 Ozone Generator

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.5.3 TEST Commands and Messages Table 5-5-17: RS-232 Test Message

Pagina 202

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 The T command lists TEST messages. Examples of the T command are: T

Pagina 203

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Whenever a warning message is reported on the analyzer display, if t

Pagina 204 - 9.3.7 Flow/Pressure Sensor

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 TABLE OF CONTENTS SAFETY MESSAGES ...

Pagina 205

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-20: RS-232 Calibration Examples Action RS-232 Commands

Pagina 206

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 5-5-21: RS-232 Calibration Messages Message Description STAR

Pagina 207

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 The following is a summary of the Diagnostic commands. Table 5-5-22:

Pagina 208

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 To print records from a DAS data channel enter: D REPORT “name” RECO

Pagina 209 - Sensor Module

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 All of the default Data Channels sample more than one parameter, for

Pagina 210

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5-40 PRINTED DOCUMENTS ARE UNCONTROLLED Variables can be changed. Be

Pagina 211

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 6 OPTIONAL HARDWARE AND SOFTWARE Optional equipment offered with th

Pagina 212

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 6-6-1: Zero/Span Valve Operation Mode Description Reference

Pagina 213

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 6.3 Internal Zero/Span (IZS) The IZS option includes the Zero/Span

Pagina 214 - 9.3.10 Pneumatic System

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 For each mode there are seven attributes that the MODE can have that

Pagina 215

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 5.2.2 CAL, CALS, CALZ, Calibration Functions ...

Pagina 216 - 10 M200A SPARE PARTS LIST

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 6-6-5: Example of AutoCal Setup Step Action Comment 1. Pre

Pagina 217 - Part No. Description

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 If an ammonia permeation tube is being used, substitute the molecula

Pagina 218

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 6-6-6: IZS Option - Permeation Tube Installation 6-7 PRINTED

Pagina 219

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 6-8 PRINTED DOCUMENTS ARE UNCONTROLLED TELEDYNE API recommends that

Pagina 220 - Includes: Qty

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7 CALIBRATION AND ZERO/SPAN CHECKS There are several ways to check

Pagina 221

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-1: Types of Zero/Span Check and Calibration Section Type

Pagina 222

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-2: Calibration Setup 7-3 PRINTED DOCUMENTS ARE UNCONTROL

Pagina 223

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.1 Manual Zero/Span Check or Cal With Zero/Span Gas in the Sample

Pagina 224

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 Table 7-7-4: Enter Expected Span Gas Concentrations Procedure Step

Pagina 225

Teledyne API Model 200A NOX Analyzer Instruction Manual, 02246, Rev. G, DCN 5247 7.2 Manual Zero/Span Check or Calibration with Zero/Span Valves Opt

Comentarios a estos manuales

Sin comentarios