INSTRUCTION MANUAL MODEL 360E CARBON DIOXIDE ANALYZER © TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION 9480 CARROLL PARK DR
M360E Documentation Model 360E Instruction Manual LIST OF APPENDICES APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION APPENDIX A-1: M360E Softwar
Operating Instructions Model 360E Instruction Manual The signals from these two connectors are routed from the motherboard via a wiring harness to tw
Model 360E Instruction Manual Operating Instructions 6.11.3. RS-485 Configuration of COM2 As delivered from the factory, COM2 is configured for RS-2
Operating Instructions Model 360E Instruction Manual When COM2 is configured for RS-485 operation the port uses the same female DB-9 connector on the
Model 360E Instruction Manual Operating Instructions 6.11.5. COMM Port Communication Modes Each of the analyzer’s serial ports can be configured to o
Operating Instructions Model 360E Instruction Manual Press the following keys to select a communication mode for a one of the COMM Ports, such as the
Model 360E Instruction Manual Operating Instructions 6.11.6. Ethernet Card Configuration When equipped with the optional Ethernet interface, the anal
Operating Instructions Model 360E Instruction Manual Table 6-12: LAN/Internet Configuration Properties PROPERTY DEFAULT STATE DESCRIPTION DHCP STA
Model 360E Instruction Manual Operating Instructions To view the above properties, press: SAMPLE ENTER SETUP PASS : 818 8
Operating Instructions Model 360E Instruction Manual 6.11.6.3. Manually Configuring the Network IP Addresses There are several circumstances when yo
Model 360E Instruction Manual Operating Instructions STEP 2: Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK addresses by pressing: Intern
Model 360E Instruction Manual M360E Documentation Figure 6-19: APICOM Remote Control Program Interface...
Operating Instructions Model 360E Instruction Manual 6.11.6.4. Changing the Analyzer’s HOSTNAME The HOSTNAME is the name by which the analyzer appea
Model 360E Instruction Manual Operating Instructions 6.11.7. Multidrop RS-232 Set Up The RS-232 multidrop consists of a printed circuit assembly that
Operating Instructions Model 360E Instruction Manual 8. If you are adding an analyzer to the end of an already existing chain, don’t forget to remov
Model 360E Instruction Manual Operating Instructions 6.11.8. COM Port Baud Rate To select the baud rate of one of the COM Ports, press: EXAMPLE S
Operating Instructions Model 360E Instruction Manual 6.11.9. COM Port Testing The serial ports can be tested for correct connection and output in the
Model 360E Instruction Manual Operating Instructions 6.12. SETUP MORE VARS: Internal Variables (VARS) The M360E has several-user adjustable softw
Operating Instructions Model 360E Instruction Manual To access and navigate the VARS menu, use the following key sequence. SETUP X.X 2 ) STABIL_
Model 360E Instruction Manual Operating Instructions 6.13. SETUP MORE DIAG: Using the Diagnostics Functions A series of diagnostic tools is gro
Operating Instructions Model 360E Instruction Manual 6.13.1. Accessing the Diagnostic Features To access the DIAG functions press the following keys:
Model 360E Instruction Manual Operating Instructions To enter the signal I/O test mode, press: SAMPLE RANGE = 500.000 PPM CO2 =X.XX
M360E Documentation Model 360E Instruction Manual x 05232 Rev B3 Table 6-3: List of Warning Messages ...
Operating Instructions Model 360E Instruction Manual 6.13.4. Analog I/O Configuration The analog I/O functions that are available in the M360E are: T
Model 360E Instruction Manual Operating Instructions The following DC current output limits apply to the current loop modules: Table 6-18: Analog O
Operating Instructions Model 360E Instruction Manual 6.13.4.1. Analog Output Signal Type and Range Span Selection To select an output signal type (DC
Model 360E Instruction Manual Operating Instructions To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (See Section 6.1
Operating Instructions Model 360E Instruction Manual To select manual output calibration for a particular channel, activate the ANALOG I/O CONFIGURAT
Model 360E Instruction Manual Operating Instructions V +DC Gnd V OUT + V OUT - V IN + V IN - RecordingDevice ANALYZER See Table 3-1 for pin
Operating Instructions Model 360E Instruction Manual 6.13.4.4. Current Loop Output Adjustment A current loop option is available and can be installed
Model 360E Instruction Manual Operating Instructions DIAG AIO CONC_OUT_2 SPAN: 10000 mV U100 UP10 UP DOWN DN10 D100 ENTR EXIT DIAG
Operating Instructions Model 360E Instruction Manual 6.13.4.5. AIN Calibration This is the sub-menu to conduct the analog input calibration. This ca
Model 360E Instruction Manual Operating Instructions 6.13.5. Electric Test The electric test function substitutes simulated signals for CO2 MEAS and
Model 360E Instruction Manual M360E Documentation 1. M360E DOCUMENTATION Thank you for purchasing the Model 300E Gas Filter Correlation Carbon monoxi
Operating Instructions Model 360E Instruction Manual 6.13.6. Dark Calibration Test The dark calibration test interrupts the signal path between the I
Model 360E Instruction Manual Operating Instructions 6.13.7. Pressure Calibration A sensor at the exit of the sample chamber continuously measures th
Operating Instructions Model 360E Instruction Manual 6.13.8. Flow Calibration The flow calibration allows the user to adjust the values of the sample
Model 360E Instruction Manual Operating Instructions 6.13.9. Test Channel Output When activated, output channel A4 can be used to report one of the t
Operating Instructions Model 360E Instruction Manual 6.14. SETUP MORE ALRM: Using the Gas Concentration Alarms The Model 360E includes two CO2 co
Model 360E Instruction Manual Operating Instructions 6.15. Remote Operation of the Analyzer 6.15.1. Remote Operation Using the External Digital I/O 6
Operating Instructions Model 360E Instruction Manual The pin assignments for the Status Outputs are: Table 6-24: Status Output Pin Assignments OUTP
Model 360E Instruction Manual Operating Instructions 6.15.1.2. Control Inputs These inputs allow the user to remotely initiate Zero and Span calibrat
Operating Instructions Model 360E Instruction Manual 6.15.2. Remote Operation Using the External Serial I/O 6.15.2.1. Terminal Operating Modes The Mo
Model 360E Instruction Manual Operating Instructions 6.15.2.3. Command Syntax Commands are not case-sensitive and all arguments within one command (i
M360E Documentation Model 360E Instruction Manual 2 05232 Rev B3 5.0 OPTIONAL HARDWARE & SOFTWARE A description of optional equipment to add fun
Operating Instructions Model 360E Instruction Manual Boolean expressions are used to specify the value of variables or I/O signals that may assume
Model 360E Instruction Manual Operating Instructions 6.15.2.6. Remote Access by Modem The M360E can be connected to a modem for remote access. This
Operating Instructions Model 360E Instruction Manual To initialize the modem press: Select which COM Port is tested SETUP X.X SECONDARY SETUP MEN
Model 360E Instruction Manual Operating Instructions 6.15.2.8. APICOM Remote Control Program APICOM is an easy-to-use, yet powerful interface program
Operating Instructions Model 360E Instruction Manual 6.15.4. Using the M360E with a Hessen Protocol Network 6.15.4.1. General Overview of Hessen Prot
Model 360E Instruction Manual Operating Instructions 6.15.4.3. Activating Hessen Protocol The first step in configuring the M360E to operate over a H
Operating Instructions Model 360E Instruction Manual NOTE While Hessen Protocol Mode can be activated independently for COM1 and COM2, The TYPE selec
Model 360E Instruction Manual Operating Instructions 6.15.4.7. Setting Hessen Protocol Status Flags Teledyne Instruments’ implementation of Hessen pr
Operating Instructions Model 360E Instruction Manual To assign or reset the status flag bit assignments, press: SETUP X.X SECONDARY SETUP MENU C
Model 360E Instruction Manual Calibration Procedures 7. CALIBRATION PROCEDURES This contains a variety of information regarding the various methods f
Model 360E Instruction Manual Specifications, Approvals and Warranty 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1. Specifications Table 2-1: Mod
Calibration Procedures Model 360E Instruction Manual 7.1.2. Calibration Gas Traceability All equipment used to produce calibration gases should be ve
Model 360E Instruction Manual Calibration Procedures VENT Figure 7-2: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator STE
Calibration Procedures Model 360E Instruction Manual STEP THREE: Perform the Zero/Span Calibration Procedure: M-P CAL STABIL=XXX.X PP
Model 360E Instruction Manual Calibration Procedures 7.3. Manual Calibration Checks Informal calibration checks, which only evaluate but do not alter
Calibration Procedures Model 360E Instruction Manual VENT Figure 7-3: Pneumatic Connections–M360E with Zero/Span/Shutoff Valves (OPT 50) So
Model 360E Instruction Manual Calibration Procedures Figure 7-5: Pneumatic Connections–M360E with Zero/Span Valves (OPT 52) VENT Figure 7-6:
Calibration Procedures Model 360E Instruction Manual STEP TWO: Set the expected CO2 Span Gas concentration: This sequence causes the analyzer to pr
Model 360E Instruction Manual Calibration Procedures STEP THREE: Perform the zero/span calibration. Zero and span checks using the zero/span valve
Calibration Procedures Model 360E Instruction Manual 7.5. Manual Calibration Checks with Zero/Span Valves Zero and span checks using the VARIOUS zero
Model 360E Instruction Manual Calibration Procedures 7.5.1. Zero/Span Calibration on Auto Range or Dual Ranges If the analyzer is being operated in d
Specifications, Approvals and Warranty Model 360E Instruction Manual 2.2. CE Mark Compliance Emissions Compliance The Teledyne Instruments Model 360E
Calibration Procedures Model 360E Instruction Manual 7.5.2. Use of Zero/Span Valves with Remote Contact Closure Contact closures for controlling cali
Model 360E Instruction Manual Calibration Procedures The following example sets Sequence #2 to do a Zero-Span Calibration every other day starting at
Calibration Procedures Model 360E Instruction Manual To program the Sequence: Toggle keys to set number of days between procedures (1-367) Toggle k
Model 360E Instruction Manual Calibration Procedures 7.6.1. AutoCal with Auto or Dual Reporting Ranges Modes Selected SETUP C.4 RANGE TO
Calibration Procedures Model 360E Instruction Manual 152 05232 Rev B3 The default iDAS configuration records all calibration values in channel CALDA
Model 360E Instruction Manual EPA Protocol Calibration 8. EPA PROTOCOL CALIBRATION At the writing of this manual there is no EPA requirements for the
Model 360E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9. MAINTENANCE SCHEDULE & PROCEDURES Predictive diagnostic functions, includi
MAINTENANCE SCHEDULE & PROCEDURES Model 360E Instruction Manual Table 9-1: M360E Maintenance Schedule DATE PERFORMED ITEM ACTION FREQ CAL CH
Model 360E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 05232 Rev B3 157 Table 9-2: M360E Test Function Record DATE RECORDED FUNCTI
Model 360E Instruction Manual Specifications, Approvals and Warranty 05232 Rev B3 5 THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREF
MAINTENANCE SCHEDULE & PROCEDURES Model 360E Instruction Manual 9.2. Predicting Failures Using the Test Functions The Test Functions can be used
Model 360E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9.3. Maintenance Procedures The following procedures are to be performed periodic
MAINTENANCE SCHEDULE & PROCEDURES Model 360E Instruction Manual 9.3.3. Performing Leak Checks Leaks are the most common cause of analyzer malfunc
Model 360E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9.3.4. Performing a Sample Flow Check CAUTION Always use a separate calibrated f
Model 360E Instruction Manual THEORY OF OPERATION 10. THEORY OF OPERATION The Model 360E Gas Filter Correlation Carbon Dioxide Analyzer is a micropro
THEORY OF OPERATION Model 360E Instruction Manual IR Source IR Beam Sample Chamber Band-Pass Filter Photo-Detector Figure 10-1: Measurement Funda
Model 360E Instruction Manual THEORY OF OPERATION IR Source Photo-Detector M R GFC Wheel IR unaffected by N2 in Measurement Cell IR IS affected by
THEORY OF OPERATION Model 360E Instruction Manual IR unaffected by N2 in Measurement Cell of the GDC Wheel and no additional CO2 in the Sample Chamb
Model 360E Instruction Manual THEORY OF OPERATION Thus, the difference in the peak heights and the resulting M/R ratio is only due to CO2 and not to
THEORY OF OPERATION Model 360E Instruction Manual 10.2. Pneumatic Operation CAUTION It is important that the sample airflow system is both leak tigh
Model 360E Instruction Manual THEORY OF OPERATION 10.2.2. Flow Rate Control To maintain a constant flow rate of the sample gas through the instrument
THEORY OF OPERATION Model 360E Instruction Manual 10.2.3. Purge Gas Pressure Control In order to ensure that all of the ambient CO2 is purged from th
Model 360E Instruction Manual THEORY OF OPERATION 10.3. Electronic Operation 10.3.1. Overview Figure 10-9 shows a block diagram of the major electron
THEORY OF OPERATION Model 360E Instruction Manual Analog Outputs Sample Flow & Pressure Sensors Status Outputs: 1 – 6 Control Inputs: 1 – 8
Model 360E Instruction Manual THEORY OF OPERATION 10.3.2. CPU The Model 360E’s CPU is a, low power (5 VDC, 0.8A max), high performance, 386-based mic
THEORY OF OPERATION Model 360E Instruction Manual 10.3.3.3. GFC Wheel A synchronous AC motor turns the GFC wheel motor. For analyzers operating on 6
Model 360E Instruction Manual THEORY OF OPERATION Reference Pulses Measurement Pulses IR Beam Pulses Segment Sensor Pulses MR Sensor Pulses Figure 1
THEORY OF OPERATION Model 360E Instruction Manual Thermo-Electric Cooler Control Circuit Sample & Hold Circuits (x4) Amplifiers Dark Switch
Model 360E Instruction Manual THEORY OF OPERATION The four sample and hold circuits are: ACTIVE WHEN: DESIGNATION IR BEAM PASSING THROUGH SEGMENT SE
Model 360E Instruction Manual Getting Started 3. GETTING STARTED 3.1. Unpacking and Initial Set Up CAUTION To avoid personal injury, always use two
THEORY OF OPERATION Model 360E Instruction Manual 10.3.4.4. Photo-Detector Temperature Control The synch/demod board also contains circuitry that con
Model 360E Instruction Manual THEORY OF OPERATION Zero/Span Valve Options Any zero/span/shutoff valve options installed in the analyzer are controlle
THEORY OF OPERATION Model 360E Instruction Manual Figure 10-14: Location of relay board Status LED’s 10.3.5.2. I2C Watch Dog Circuitry Special circui
Model 360E Instruction Manual THEORY OF OPERATION 10.3.6.3. Thermistor Interface This circuit provides excitation, termination and signal selection f
THEORY OF OPERATION Model 360E Instruction Manual All four analog outputs are connected back to the A/D converter through a Loop-back circuit. This
Model 360E Instruction Manual THEORY OF OPERATION 10.3.8. Power Supply/ Circuit Breaker The analyzer operates on 100 VAC, 115 VAC or 230 VAC power at
THEORY OF OPERATION Model 360E Instruction Manual 10.4. Interface The analyzer has several ways to communicate the outside world, see Figure 10-16.
Model 360E Instruction Manual THEORY OF OPERATION The most commonly used method for communicating with the M360E Analyzer is via the instrument’s fr
THEORY OF OPERATION Model 360E Instruction Manual 10.4.1.4. Keyboard/Display Interface Electronics FRONT PANEL Keypad Decoder Key Press Dete
Model 360E Instruction Manual THEORY OF OPERATION I2C INTERFACE CHIP This IC performs several functions: Using a dedicated digital status bit, it
THIS PAGE IS INTENTIONALLY LEFT BLANK
Getting Started Model 360E Instruction Manual 5. Check the connectors of the various internal wiring harnesses and pneumatic hoses to make sure they
THEORY OF OPERATION Model 360E Instruction Manual 10.5. Software Operation The Model 360E Gas Filter Correlation Carbon Dioxide Analyzer is at its he
Model 360E Instruction Manual THEORY OF OPERATION Two conditions must be simultaneously met to switch to the short filter. First the instantaneous c
THEORY OF OPERATION Model 360E Instruction Manual 190 05232 Rev B3 10.5.5. Internal Data Acquisition System (iDAS) The iDAS is designed to implemen
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11. TROUBLESHOOTING & REPAIR PROCEDURES This contains a variety of methods
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.1.1. Interpreting WARNING Messages The most common and/or serious instrument
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES Table 11-1: Warning Messages - Indicated Failures WARNING MESSAGE FAULT CONDI
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual Table 11-1: Warning Messages – Indicated Failures (cont.) WARNING MESSAGE FAU
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES Table 11-2: Test Functions - Indicated Failures TEST FUNCTIONS INDICATED FAI
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.1.3. Using the Diagnostic Signal I/O Function The Signal I/O parameters foun
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.1.4. Internal Electronic Status LED’s Several LED’s are located inside the i
Model 360E Instruction Manual Getting Started 3.1.1. M360 Layout Figure 3-1 shows the front panel layout of the analyzer. Figure 3-2 shows the rear
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual D1 – M/R Sensor Status D2 – Segment Sensor Status JP4 Connector to Opto-Pickup
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES Table 11-5: Relay Board Status LED Failure Indications SIGNAL I/O PARAMETER L
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.2. Gas Flow Problems The M360E has two main gas flow path, the sample gas fl
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES GFC WheelHousing SAMPLE CHAMBERGFC Motor Heat SyncPurge GasFlow Contro
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual Figure 11.9: M360E – Internal Pneumatics with O2 Sensor Option 65 11.2.2. Ty
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 4. If gas flows through the instrument when it is disconnected from its source
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.2.2.6. Sample Pump The sample pump should start immediately after the front
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.3.2. Non-Repeatable Zero and Span As stated earlier, leaks both in the M360E
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.4. Other Performance Problems Dynamic problems (i.e. problems which only ma
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 4. If K2 checks out OK, the thermistor temperature sensor located on the optic
Getting Started Model 360E Instruction Manual Particulate Filter SYNC/DEMOD BOARD GFC Motor On/Off Switch & Circuit Breaker PS1 (+5 VDC; 15VD
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.4.2. Excessive Noise Noise is continuously monitored in the TEST functions a
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.5. Subsystem Checkout The preceding s of this manual discussed a variety of
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual A voltmeter should be used to verify that the DC voltages are correct per the v
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.5.5. Relay Board The relay board PCA (04135) can be most easily checked by o
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.5.6.2. Opto Pickup Assembly Operation of the opto pickup PCA (04088) can be
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 1. For Pressure related problems: Measure the voltage across C1 it should b
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual Table 11-10: Analog Output Test Function - Nominal Values Voltage Outputs F
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.5.7.4. Status Outputs The procedure below can be used to test the Status out
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.5.8. CPU There are two major types of failures associated with the CPU board
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.5.9.2. Troubleshooting Analyzer/Modem or Terminal Operation These are the ge
Model 360E Instruction Manual Getting Started GFC Temperature Sensor Shock Absorbing Mounting Bracket Purge Gas Inlet IR Source GFC Wheel Heat Sync
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.6. Repair Procedures This contains procedures that might need to be performe
Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.6.2. Removing/Replacing the GFC Wheel When removing or replacing the GFC Wh
TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11 12 Figure 11-12: Removing the GFC Wheel 12. Remove the GFC wheel/mask asse
Model 360E Instruction Manual A Primer on Electro-Static Discharge 12. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne Instruments considers the preve
A Primer on Electro-Static Discharge Model 360E Instruction Manual 12.2. How Electro-Static Charges Cause Damage Damage to components occurs when the
Model 360E Instruction Manual A Primer on Electro-Static Discharge 12.3. Common Myths About ESD Damage I didn’t feel a shock so there was no elec
A Primer on Electro-Static Discharge Model 360E Instruction Manual For technicians that work in the field, special lightweight and portable anti-ESD
Model 360E Instruction Manual A Primer on Electro-Static Discharge 12.4.2. Basic anti-ESD Procedures for Analyzer Repair and Maintenance 12.4.2.1. W
A Primer on Electro-Static Discharge Model 360E Instruction Manual 12.4.2.3. Transferring Components from Rack to Bench and Back When transferring a
Model 360E Instruction Manual A Primer on Electro-Static Discharge 12.4.2.5. Packing Components for Return to Teledyne Instruments Customer Service.
Getting Started Model 360E Instruction Manual 3.1.2. Electrical Connections 3.1.2.1. Power Connection Attach the power cord to the analyzer and plug
A Primer on Electro-Static Discharge Model 360E Instruction Manual 228 05232 Rev B3 USER NOTES:
Model 360E Instruction Manual APPENDIX A - Version Specific Software Documentation APPENDIX A - Version Specific Software Documentation APPENDIX A-1:
APPENDIX A-1: M360E Software Menu Trees, Revision G.4 Model 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 CO25 O25 Z
Model 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 SAMPLE <TST TST> SETUP COMM DIAG VARS (Secondary Setup Me
APPENDIX A-1: M360E Software Menu Trees, Revision G.4 Model 360E Instruction Manual • MODEL NAME • SERIAL NUMBER • SOFTWARE REVISION • LIBRARY R
Model 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 NAME EVENT PARAMETERS REPORT PERIOD NUMBER OF RECORDS RS-232 REP
APPENDIX A-1: M360E Software Menu Trees, Revision G.4 Model 360E Instruction Manual CFG ACAL1 SETUP CLK PASS RNGE DAS MORE COMM VARS DIAG Go To DI
Model 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 INSTRUMENT IP5 GATEWAY IP5 SUBNET MASK5 TCP PORT3 DHCP INSTRUME
APPENDIX A-1: M360E Software Menu Trees, Revision G.4 Model 360E Instruction Manual . See Fig A-5 QUIET COMPUTER SECURITY HESSEN PROTOCOL COM[1,2] E
odel 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 3 Rev G.4 A-9 CFG ACAL1 CLK PASS RNGE DAS MORE COMM VARS ANALOG
Model 360E Instruction Manual Getting Started 3.1.2.2. Output Connections The M360E is equipped with several analog output channels accessible throug
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 T
Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
Getting Started Model 360E Instruction Manual 3.1.2.3. Connecting the Status Outputs If you wish utilize the analyzer’s status outputs to interface w
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC
Model 360E Instruction Manual APPENDIX A-3: Warnings and Test Functions, Revision G.4 APPENDIX A-3: Warnings and Test Functions, Revision G.4 Table A
APPENDIX A-3: Warnings and Test Functions, Revision G.4 Model 360E Instruction Manual Table A-3: M360E Test Functions, Revision G.4 TEST FUNCTION
Model 360E Instruction Manual APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4 APPENDIX A-4: Signal I/O Definitions f
APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4 Model 360E Instruction Manual SIGNAL NAME BIT OR CHANNEL NUMBER DESCRI
Model 360E Instruction Manual APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4 SIGNAL NAME BIT OR CHANNEL NUMBER DESCRI
APPENDIX A-5: M360E iDAS Parameters Revision G.4 Model 360E Instruction Manual APPENDIX A-5: M360E iDAS Parameters Revision G.4 Table A-5: M360E
Model 360E Instruction Manual APPENDIX A-5: M360E iDAS Parameters Revision G.4 Table A-6: M360E iDAS Functions, Revision G.4 NAME DESCRIPTION U
Model 360E Instruction Manual Getting Started 3.1.2.4. Connecting the Control Inputs If you wish to use the analyzer to remotely activate the zero an
APPENDIX A-6: Terminal Command Designators, Revision G.4 Model 360E Instruction Manual APPENDIX A-6: Terminal Command Designators, Revision G.4 Table
Model 360E Instruction Manual APPENDIX A-6: Terminal Command Designators, Revision G.4 Table A-8: Terminal Key Assignments, Revision G.4 TERMINAL
APPENDIX A-6: Terminal Command Designators, Revision G.4 Model 360E Instruction Manual USER NOTES: A-32 05233 Rev G.4
Model 360E Instruction Manual APPENDIX B - M360E Spare Parts List 05234 Rev B B-1 APPENDIX B - M360E Spare Parts List NOTE Use of replacement parts
APPENDIX B - M360E Spare Parts List Model 360E Instruction Manual B-2 05234 Rev B
M360E Spare Parts ListPart NumberDescription000941000 ORIFICE, 13 MIL (SAMPLE FLOW)001760400 ASSY, FLOW CTL, 800CC, 1/
M360E Spare Parts ListPart NumberDescription050320000 PCA, OPTO-INTERRUPTER, M300E 052560000 PCA, OPTO-INTE
Recommended Spare Parts Stocking Levels Model 360E1 2-5 6-10 11-20 21-30003290500 ASSY, THERMISTOR, M300 BENCH 1 1 1 1 200
THIS PAGE IS INTENTIONALLY LEFT BLANK
Model 360E Instruction Manual Warranty/Repair Questionnaire TELEDYNE INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies C
Getting Started Model 360E Instruction Manual 3.1.2.5. Connecting the Serial Ports If you wish to utilize either of the analyzer’s two serial interfa
Model 360E Instruction Manual Warranty/Repair Questionnaire TELEDYNE INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies C
Model360E Instruction Manual APPENDIX D - ELECTRONIC SCHEMATICS 05236 Rev C D-1 APPENDIX D - ELECTRONIC SCHEMATICS Table D-1: List of Included
APPENDIX D - ELECTRONIC SCHEMATICS Model 360E Instruction Manual D-2 05236 Rev C
Model 360E Instruction Manual Getting Started Table 3-4: Model 360E Rear Panel Pneumatic Connections REAR PANEL LABEL FUNCTION SAMPLE Connect a
Model 360E Instruction Manual M360E Documentation SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many imp
Getting Started Model 360E Instruction Manual VENT Figure 3-7: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator 1. Attac
Model 360E Instruction Manual Getting Started EXAMPLE: If the application is to measure between 0 ppm and 500 ppm, an appropriate Span Gas would be
Getting Started Model 360E Instruction Manual 3.1.3.2. Connections with Internal Valve Options Installed Figures 3-8 through 3-11 show the proper pne
Model 360E Instruction Manual Getting Started Figure 3-10: Pneumatic Connections–M360E with Zero/Span Valves (OPT 52) Source ofSAMPLE GASRemove
Getting Started Model 360E Instruction Manual 3.1.3.3. Pneumatic Connections to M360E in Multipoint Calibration Applications Some applications may re
Model 360E Instruction Manual Getting Started 3.2. Initial Operation If you are unfamiliar with the M360E theory of operation, we recommend that you
Getting Started Model 360E Instruction Manual 3.2.2. Warm Up The M360E requires about 30 minutes warm-up time before reliable CO2 measurements can be
Model 360E Instruction Manual Getting Started Table 3-6: Possible Warning Messages at Start-Up MESSAGE MEANING ANALOG CAL WARNING The instrument’
Getting Started Model 360E Instruction Manual 3.2.4. Functional Check 1. After the analyzer’s components has warmed up for at least 30 minutes, veri
Model 360E Instruction Manual Getting Started 3.3. Initial Calibration Procedure The next task is to calibrate the analyzer. To perform the followi
Getting Started Model 360E Instruction Manual 2. Set the expected CO2 span gas concentration This sequence causes the analyzer to prompt for the e
Model 360E Instruction Manual Getting Started 3. Perform the Zero/Span Calibration Procedure M-P CAL STABIL=XXX.X PPM CO2 =XXX
Getting Started Model 360E Instruction Manual 3.3.1. Initial O2 Sensor Calibration Procedure If your instrument has an O2 sensor option installed tha
Model 360E Instruction Manual Getting Started STEP 2 – ACTIVATE O2 SENSOR STABILITY FUNCTION To change the stability test function from NOx concentra
Getting Started Model 360E Instruction Manual STEP 4 – O2 Zero/Span Calibration: To perform the zero/span calibration procedure: 32 05232 Rev B3
Model 360E Instruction Manual Getting Started 05232 Rev B3 33 The Model 360E Analyzer is now ready for operation NOTE Once you have completed the a
Model 360E Instruction Manual Frequently Asked Questions 4. FREQUENTLY ASKED QUESTIONS 4.1. FAQ’s The following is a list from the Teledyne Instrumen
Frequently Asked Questions Model 360E Instruction Manual A: These components are located inside the flow control assembly that is attached to the inl
Model 360E Instruction Manual Frequently Asked Questions 05232 Rev B3 37 IC – Acronym for Integrated Circuit, a modern, semi-conductor circuit that
Model 360E Instruction Manual M360E Documentation TABLE OF CONTENTS 1. M360E DOCUMENTATION...
Model 360E Instruction Manual Optional Hardware and Software 5. OPTIONAL HARDWARE AND SOFTWARE This includes a brief description of the hardware and
Optional Hardware and Software Model 360E Instruction Manual 5.2.1. Converting Current Loop Analog Outputs to Standard Voltage Outputs. NOTE Servicin
Model 360E Instruction Manual Optional Hardware and Software 5.4. Calibration Valves Options There are four available options involving Zero/Span/Shu
Optional Hardware and Software Model 360E Instruction Manual GFC WheelHousing SAMPLE CHAMBERGFC Motor Heat SyncPurge GasFlow ControlExternal
Model 360E Instruction Manual Optional Hardware and Software Table 5-2: Zero/Span Valve Operating States for Options 52 & 53 Mode Valve Cond
Optional Hardware and Software Model 360E Instruction Manual 5.4.4. Zero/Span Valve with External CO2 Scrubber (Option 53) Option 53 is operationally
Model 360E Instruction Manual Optional Hardware and Software 5.5.3. Ethernet (Option 63) When installed, this option is electronically connected to t
Optional Hardware and Software Model 360E Instruction Manual 5.6. Oxygen Sensor (OPT 65) 5.6.1. Theory of Operation 5.6.1.1. Paramagnetic measurement
Model 360E Instruction Manual Optional Hardware and Software The O2 concentration range is 0-100% (user selectable) with 0.1% precision and accuracy
M360E Documentation Model 360E Instruction Manual 6. OPERATING INSTRUCTIONS...
Optional Hardware and Software Model 360E Instruction Manual 5.7. Additional Manuals 5.7.1. Printed Manuals (Option 70) Additional printed copies of
Model 360E Instruction Manual Optional Hardware and Software 05232 Rev B3 49 5.9.2. Maintenance Mode Switch API’s instruments can be equipped with a
Model 360E Instruction Manual Operating Instructions 6. OPERATING INSTRUCTIONS To assist in navigating the analyzer’s software, a series of menu tree
Operating Instructions Model 360E Instruction Manual Table 6-1: Analyzer Operating modes MODE MEANING DIAG One of the analyzer’s diagnostic modes
Model 360E Instruction Manual Operating Instructions Table 6-2: Test Functions Defined PARAMETER DISPLAY TITLE UNITS MEANING Range RANGE RANGE11
Operating Instructions Model 360E Instruction Manual RANGERANGE11 RANGE21 02 RANGE STABIL MEAS REF MR RATIO PRES SAMP FL SAMP TEMP BENCH TEMPWHEEL T
Model 360E Instruction Manual Operating Instructions 6.2.2. Warning Messages The most common instrument failures will be reported as a warning on the
Operating Instructions Model 360E Instruction Manual To view and clear warning messages SAMPLE HVPS WARNING CO2 = 0.00
Model 360E Instruction Manual Operating Instructions 6.4. SETUP MODE The SETUP mode contains a variety of choices that are used to configure the anal
Model 360E Instruction Manual M360E Documentation 6.13.8. Flow Calibration...
Operating Instructions Model 360E Instruction Manual 6.5. SETUP CFG: Viewing the Analyzer’s Configuration Information Pressing the CFG key displays
Model 360E Instruction Manual Operating Instructions 6.7. SETUP DAS: Using the Data Acquisition System (iDAS) The M360E analyzer contains a flexibl
Operating Instructions Model 360E Instruction Manual 6.7.1.1. iDAS Channels The key to the flexibility of the iDAS is its ability to store a large nu
Model 360E Instruction Manual Operating Instructions 6.7.1.2. iDAS Parameters Data parameters are types of data that may be measured and stored by th
Operating Instructions Model 360E Instruction Manual 6.7.1.3. iDAS Triggering Events Triggering events define when and how the iDAS records a measure
Model 360E Instruction Manual Operating Instructions Note The CALDAT, STBZRO and STBSPN channels collect data based on events (e.g. a calibration op
Operating Instructions Model 360E Instruction Manual NOTE Sending an iDAS configuration to the analyzer through its COM ports will replace the existi
Model 360E Instruction Manual Operating Instructions 6.7.2.1. Viewing iDAS Data and Settings iDAS data and settings can be viewed on the front panel
Operating Instructions Model 360E Instruction Manual 6.7.2.2. Editing iDAS Data Channels iDAS configuration is most conveniently done through the API
Model 360E Instruction Manual Operating Instructions From the end of the previous key sequence … SETUP X.X NAME:CONC C O
M360E Documentation Model 360E Instruction Manual 10.2.1. Sample Gas Flow ...
Operating Instructions Model 360E Instruction Manual 6.7.2.4. Editing iDAS Parameters Data channels can be edited individually from the front panel w
Model 360E Instruction Manual Operating Instructions To configure a specific data parameter, press: FROM THE EDIT DATA PARAMETER MENU (see previo
Operating Instructions Model 360E Instruction Manual 6.7.2.5. Sample Period and Report Period The iDAS defines two principal time periods by which sa
Model 360E Instruction Manual Operating Instructions To define the REPORT PERIOD, follow the instruction shown in Section 6.7.2.2 then press: Edit D
Operating Instructions Model 360E Instruction Manual 6.7.2.6. Number of Records The number of data records in the M360E is limited to about a cumulat
Model 360E Instruction Manual Operating Instructions 6.7.2.7. RS-232 Report Function The M360E iDAS can automatically report data to the communicatio
Operating Instructions Model 360E Instruction Manual 6.7.2.10. Disabling/Enabling Data Channels Data channels can be temporarily disabled, which can
Model 360E Instruction Manual Operating Instructions 6.7.3. Remote iDAS Configuration Editing channels, parameters and triggering events as described
Operating Instructions Model 360E Instruction Manual Figure 6-6: iDAS Configuration Through a Terminal Emulation Program. Both procedures are be
Model 360E Instruction Manual Operating Instructions 6.8. SETUP RNGE: Analog Output Reporting Range Configuration The analyzer has three active ana
Model 360E Instruction Manual M360E Documentation 11.2.3. Poor or Stopped Flow of Purge Gas...
Operating Instructions Model 360E Instruction Manual The M360E solves this problem by allowing the user to select a scaled reporting range for the an
Model 360E Instruction Manual Operating Instructions NOTE Upper span limit setting for the individual range modes are shared. Resetting the span lim
Operating Instructions Model 360E Instruction Manual 6.8.4. Dual Range Mode (DUAL) Selecting Dual Range mode allows the A1 and A2 outputs to be confi
Model 360E Instruction Manual Operating Instructions 6.8.5. Auto Range Mode (AUTO) In AUTO range mode, the analyzer automatically switches the report
Operating Instructions Model 360E Instruction Manual 6.8.6. Range Units The M360E can display concentrations in parts per billion (109 mols per mol,
Model 360E Instruction Manual Operating Instructions 6.8.7. Dilution Ratio The dilution ratio is a software option that allows the user to compensate
Operating Instructions Model 360E Instruction Manual 6.9. SETUP PASS: Password Feature The M360E provides password protection of the calibration an
Model 360E Instruction Manual Operating Instructions 6.10. SETUP CLK: Setting the Internal Time-of-Day Clock The M360E has a time of day clock tha
Operating Instructions Model 360E Instruction Manual 6.11. SETUP MORE COMM: Using the Analyser’s Communication Ports The M360E is equipped with tw
Model 360E Instruction Manual Operating Instructions 6.11.2. COMM Port Default Settings Received from the factory, the analyzer is set up to emulate