Pagina 1 - CARBON DIOXIDE ANALYZER

INSTRUCTION MANUAL MODEL 360E CARBON DIOXIDE ANALYZER © TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION 9480 CARROLL PARK DR

Pagina 2

M360E Documentation Model 360E Instruction Manual LIST OF APPENDICES APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION APPENDIX A-1: M360E Softwar

Pagina 3 - SAFETY MESSAGES

Operating Instructions Model 360E Instruction Manual The signals from these two connectors are routed from the motherboard via a wiring harness to tw

Pagina 4

Model 360E Instruction Manual Operating Instructions 6.11.3. RS-485 Configuration of COM2 As delivered from the factory, COM2 is configured for RS-2

Pagina 5 - TABLE OF CONTENTS

Operating Instructions Model 360E Instruction Manual When COM2 is configured for RS-485 operation the port uses the same female DB-9 connector on the

Pagina 6

Model 360E Instruction Manual Operating Instructions 6.11.5. COMM Port Communication Modes Each of the analyzer’s serial ports can be configured to o

Pagina 7 - 2 Rev B3 v

Operating Instructions Model 360E Instruction Manual Press the following keys to select a communication mode for a one of the COMM Ports, such as the

Pagina 8

Model 360E Instruction Manual Operating Instructions 6.11.6. Ethernet Card Configuration When equipped with the optional Ethernet interface, the anal

Pagina 9 - 2 Rev B3 vii

Operating Instructions Model 360E Instruction Manual Table 6-12: LAN/Internet Configuration Properties PROPERTY DEFAULT STATE DESCRIPTION DHCP STA

Pagina 10 - LIST OF FIGURES

Model 360E Instruction Manual Operating Instructions To view the above properties, press: SAMPLE ENTER SETUP PASS : 818 8

Pagina 11 - LIST OF TABLES

Operating Instructions Model 360E Instruction Manual 6.11.6.3. Manually Configuring the Network IP Addresses There are several circumstances when yo

Pagina 12 - User Notes

Model 360E Instruction Manual Operating Instructions STEP 2: Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK addresses by pressing: Intern

Pagina 13 - 1. M360E DOCUMENTATION

Model 360E Instruction Manual M360E Documentation Figure 6-19: APICOM Remote Control Program Interface...

Pagina 14

Operating Instructions Model 360E Instruction Manual 6.11.6.4. Changing the Analyzer’s HOSTNAME The HOSTNAME is the name by which the analyzer appea

Pagina 15 - 2.1. Specifications

Model 360E Instruction Manual Operating Instructions 6.11.7. Multidrop RS-232 Set Up The RS-232 multidrop consists of a printed circuit assembly that

Pagina 16 - 2.3. Warranty

Operating Instructions Model 360E Instruction Manual 8. If you are adding an analyzer to the end of an already existing chain, don’t forget to remov

Pagina 17

Model 360E Instruction Manual Operating Instructions 6.11.8. COM Port Baud Rate To select the baud rate of one of the COM Ports, press: EXAMPLE S

Pagina 18

Operating Instructions Model 360E Instruction Manual 6.11.9. COM Port Testing The serial ports can be tested for correct connection and output in the

Pagina 19 - 3. GETTING STARTED

Model 360E Instruction Manual Operating Instructions 6.12. SETUP  MORE  VARS: Internal Variables (VARS) The M360E has several-user adjustable softw

Pagina 20 - Back of the instrument

Operating Instructions Model 360E Instruction Manual To access and navigate the VARS menu, use the following key sequence. SETUP X.X 2 ) STABIL_

Pagina 21 - 3.1.1. M360 Layout

Model 360E Instruction Manual Operating Instructions 6.13. SETUP  MORE  DIAG: Using the Diagnostics Functions A series of diagnostic tools is gro

Pagina 22 - (+12 VDC)

Operating Instructions Model 360E Instruction Manual 6.13.1. Accessing the Diagnostic Features To access the DIAG functions press the following keys:

Pagina 23

Model 360E Instruction Manual Operating Instructions To enter the signal I/O test mode, press: SAMPLE RANGE = 500.000 PPM CO2 =X.XX

Pagina 24 - 3.1.2.1. Power Connection

M360E Documentation Model 360E Instruction Manual x 05232 Rev B3 Table 6-3: List of Warning Messages ...

Pagina 25 - 3.1.2.2. Output Connections

Operating Instructions Model 360E Instruction Manual 6.13.4. Analog I/O Configuration The analog I/O functions that are available in the M360E are: T

Pagina 26

Model 360E Instruction Manual Operating Instructions The following DC current output limits apply to the current loop modules: Table 6-18: Analog O

Pagina 27

Operating Instructions Model 360E Instruction Manual 6.13.4.1. Analog Output Signal Type and Range Span Selection To select an output signal type (DC

Pagina 28

Model 360E Instruction Manual Operating Instructions To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (See Section 6.1

Pagina 29

Operating Instructions Model 360E Instruction Manual To select manual output calibration for a particular channel, activate the ANALOG I/O CONFIGURAT

Pagina 30

Model 360E Instruction Manual Operating Instructions V +DC Gnd V OUT + V OUT - V IN + V IN - RecordingDevice ANALYZER See Table 3-1 for pin

Pagina 31

Operating Instructions Model 360E Instruction Manual 6.13.4.4. Current Loop Output Adjustment A current loop option is available and can be installed

Pagina 32

Model 360E Instruction Manual Operating Instructions DIAG AIO CONC_OUT_2 SPAN: 10000 mV U100 UP10 UP DOWN DN10 D100 ENTR EXIT DIAG

Pagina 33

Operating Instructions Model 360E Instruction Manual 6.13.4.5. AIN Calibration This is the sub-menu to conduct the analog input calibration. This ca

Pagina 34

Model 360E Instruction Manual Operating Instructions 6.13.5. Electric Test The electric test function substitutes simulated signals for CO2 MEAS and

Pagina 35 - 3.2. Initial Operation

Model 360E Instruction Manual M360E Documentation 1. M360E DOCUMENTATION Thank you for purchasing the Model 300E Gas Filter Correlation Carbon monoxi

Pagina 36 - 3.2.3. Warning Messages

Operating Instructions Model 360E Instruction Manual 6.13.6. Dark Calibration Test The dark calibration test interrupts the signal path between the I

Pagina 37

Model 360E Instruction Manual Operating Instructions 6.13.7. Pressure Calibration A sensor at the exit of the sample chamber continuously measures th

Pagina 38 - 3.2.4. Functional Check

Operating Instructions Model 360E Instruction Manual 6.13.8. Flow Calibration The flow calibration allows the user to adjust the values of the sample

Pagina 39

Model 360E Instruction Manual Operating Instructions 6.13.9. Test Channel Output When activated, output channel A4 can be used to report one of the t

Pagina 40

Operating Instructions Model 360E Instruction Manual 6.14. SETUP MORE  ALRM: Using the Gas Concentration Alarms The Model 360E includes two CO2 co

Pagina 41 - 05232 Rev B3 29

Model 360E Instruction Manual Operating Instructions 6.15. Remote Operation of the Analyzer 6.15.1. Remote Operation Using the External Digital I/O 6

Pagina 42 - 3.3.1. Initial O

Operating Instructions Model 360E Instruction Manual The pin assignments for the Status Outputs are: Table 6-24: Status Output Pin Assignments OUTP

Pagina 43

Model 360E Instruction Manual Operating Instructions 6.15.1.2. Control Inputs These inputs allow the user to remotely initiate Zero and Span calibrat

Pagina 44

Operating Instructions Model 360E Instruction Manual 6.15.2. Remote Operation Using the External Serial I/O 6.15.2.1. Terminal Operating Modes The Mo

Pagina 45

Model 360E Instruction Manual Operating Instructions 6.15.2.3. Command Syntax Commands are not case-sensitive and all arguments within one command (i

Pagina 46

M360E Documentation Model 360E Instruction Manual 2 05232 Rev B3 5.0 OPTIONAL HARDWARE & SOFTWARE A description of optional equipment to add fun

Pagina 47 - 4.1. FAQ’s

Operating Instructions Model 360E Instruction Manual  Boolean expressions are used to specify the value of variables or I/O signals that may assume

Pagina 48 - 4.2. Glossary

Model 360E Instruction Manual Operating Instructions 6.15.2.6. Remote Access by Modem The M360E can be connected to a modem for remote access. This

Pagina 49

Operating Instructions Model 360E Instruction Manual To initialize the modem press: Select which COM Port is tested SETUP X.X SECONDARY SETUP MEN

Pagina 50

Model 360E Instruction Manual Operating Instructions 6.15.2.8. APICOM Remote Control Program APICOM is an easy-to-use, yet powerful interface program

Pagina 51 - OPTION NUMBER DESCRIPTION

Operating Instructions Model 360E Instruction Manual 6.15.4. Using the M360E with a Hessen Protocol Network 6.15.4.1. General Overview of Hessen Prot

Pagina 52 - Outputs

Model 360E Instruction Manual Operating Instructions 6.15.4.3. Activating Hessen Protocol The first step in configuring the M360E to operate over a H

Pagina 53 - (Normal State)

Operating Instructions Model 360E Instruction Manual NOTE While Hessen Protocol Mode can be activated independently for COM1 and COM2, The TYPE selec

Pagina 54

Model 360E Instruction Manual Operating Instructions 6.15.4.7. Setting Hessen Protocol Status Flags Teledyne Instruments’ implementation of Hessen pr

Pagina 55

Operating Instructions Model 360E Instruction Manual To assign or reset the status flag bit assignments, press: SETUP X.X SECONDARY SETUP MENU C

Pagina 56 - 5.5. Communication Options

Model 360E Instruction Manual Calibration Procedures 7. CALIBRATION PROCEDURES This contains a variety of information regarding the various methods f

Pagina 57 - 5.5.3. Ethernet (Option 63)

Model 360E Instruction Manual Specifications, Approvals and Warranty 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1. Specifications Table 2-1: Mod

Pagina 58 - 5.6. Oxygen Sensor (OPT 65)

Calibration Procedures Model 360E Instruction Manual 7.1.2. Calibration Gas Traceability All equipment used to produce calibration gases should be ve

Pagina 59 - Sensor

Model 360E Instruction Manual Calibration Procedures VENT Figure 7-2: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator STE

Pagina 60 - 5.7. Additional Manuals

Calibration Procedures Model 360E Instruction Manual STEP THREE: Perform the Zero/Span Calibration Procedure: M-P CAL STABIL=XXX.X PP

Pagina 61

Model 360E Instruction Manual Calibration Procedures 7.3. Manual Calibration Checks Informal calibration checks, which only evaluate but do not alter

Pagina 62

Calibration Procedures Model 360E Instruction Manual VENT Figure 7-3: Pneumatic Connections–M360E with Zero/Span/Shutoff Valves (OPT 50) So

Pagina 63 - 6. OPERATING INSTRUCTIONS

Model 360E Instruction Manual Calibration Procedures Figure 7-5: Pneumatic Connections–M360E with Zero/Span Valves (OPT 52) VENT Figure 7-6:

Pagina 64 - 6.2. Sample Mode

Calibration Procedures Model 360E Instruction Manual STEP TWO: Set the expected CO2 Span Gas concentration: This sequence causes the analyzer to pr

Pagina 65

Model 360E Instruction Manual Calibration Procedures STEP THREE: Perform the zero/span calibration. Zero and span checks using the zero/span valve

Pagina 66

Calibration Procedures Model 360E Instruction Manual 7.5. Manual Calibration Checks with Zero/Span Valves Zero and span checks using the VARIOUS zero

Pagina 67 - 6.2.2. Warning Messages

Model 360E Instruction Manual Calibration Procedures 7.5.1. Zero/Span Calibration on Auto Range or Dual Ranges If the analyzer is being operated in d

Pagina 68 - 6.3. Calibration Mode

Specifications, Approvals and Warranty Model 360E Instruction Manual 2.2. CE Mark Compliance Emissions Compliance The Teledyne Instruments Model 360E

Pagina 69 - 6.4. SETUP MODE

Calibration Procedures Model 360E Instruction Manual 7.5.2. Use of Zero/Span Valves with Remote Contact Closure Contact closures for controlling cali

Pagina 70 - Information

Model 360E Instruction Manual Calibration Procedures The following example sets Sequence #2 to do a Zero-Span Calibration every other day starting at

Pagina 71 - 6.7.1. iDAS Structure

Calibration Procedures Model 360E Instruction Manual To program the Sequence: Toggle keys to set number of days between procedures (1-367) Toggle k

Pagina 72 - 6.7.1.1. iDAS Channels

Model 360E Instruction Manual Calibration Procedures 7.6.1. AutoCal with Auto or Dual Reporting Ranges Modes Selected SETUP C.4 RANGE TO

Pagina 73 - 6.7.1.2. iDAS Parameters

Calibration Procedures Model 360E Instruction Manual 152 05232 Rev B3 The default iDAS configuration records all calibration values in channel CALDA

Pagina 74 - 6.7.2. Default iDAS Channels

Model 360E Instruction Manual EPA Protocol Calibration 8. EPA PROTOCOL CALIBRATION At the writing of this manual there is no EPA requirements for the

Pagina 76

Model 360E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9. MAINTENANCE SCHEDULE & PROCEDURES Predictive diagnostic functions, includi

Pagina 77 - 05232 Rev B3 65

MAINTENANCE SCHEDULE & PROCEDURES Model 360E Instruction Manual Table 9-1: M360E Maintenance Schedule DATE PERFORMED ITEM ACTION FREQ CAL CH

Pagina 78 - DATA ACQUISITION

Model 360E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 05232 Rev B3 157 Table 9-2: M360E Test Function Record DATE RECORDED FUNCTI

Pagina 79 - 6.7.2.3. Trigger Events

Model 360E Instruction Manual Specifications, Approvals and Warranty 05232 Rev B3 5 THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREF

Pagina 80 - Press SET> key until…

MAINTENANCE SCHEDULE & PROCEDURES Model 360E Instruction Manual 9.2. Predicting Failures Using the Test Functions The Test Functions can be used

Pagina 81 - 05232 Rev B3 69

Model 360E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9.3. Maintenance Procedures The following procedures are to be performed periodic

Pagina 82

MAINTENANCE SCHEDULE & PROCEDURES Model 360E Instruction Manual 9.3.3. Performing Leak Checks Leaks are the most common cause of analyzer malfunc

Pagina 83 - 05232 Rev B3 71

Model 360E Instruction Manual MAINTENANCE SCHEDULE & PROCEDURES 9.3.4. Performing a Sample Flow Check CAUTION Always use a separate calibrated f

Pagina 85

Model 360E Instruction Manual THEORY OF OPERATION 10. THEORY OF OPERATION The Model 360E Gas Filter Correlation Carbon Dioxide Analyzer is a micropro

Pagina 86

THEORY OF OPERATION Model 360E Instruction Manual IR Source IR Beam Sample Chamber Band-Pass Filter Photo-Detector Figure 10-1: Measurement Funda

Pagina 87

Model 360E Instruction Manual THEORY OF OPERATION IR Source Photo-Detector M R GFC Wheel IR unaffected by N2 in Measurement Cell IR IS affected by

Pagina 88

THEORY OF OPERATION Model 360E Instruction Manual IR unaffected by N2 in Measurement Cell of the GDC Wheel and no additional CO2 in the Sample Chamb

Pagina 89 - Configuration

Model 360E Instruction Manual THEORY OF OPERATION Thus, the difference in the peak heights and the resulting M/R ratio is only due to CO2 and not to

Pagina 91 - PRIMARY SETUP MENU

THEORY OF OPERATION Model 360E Instruction Manual 10.2. Pneumatic Operation CAUTION It is important that the sample airflow system is both leak tigh

Pagina 92

Model 360E Instruction Manual THEORY OF OPERATION 10.2.2. Flow Rate Control To maintain a constant flow rate of the sample gas through the instrument

Pagina 93 - PRIMARY SETUP MENU

THEORY OF OPERATION Model 360E Instruction Manual 10.2.3. Purge Gas Pressure Control In order to ensure that all of the ambient CO2 is purged from th

Pagina 94 - 6.8.6. Range Units

Model 360E Instruction Manual THEORY OF OPERATION 10.3. Electronic Operation 10.3.1. Overview Figure 10-9 shows a block diagram of the major electron

Pagina 95 - 6.8.7. Dilution Ratio

THEORY OF OPERATION Model 360E Instruction Manual Analog Outputs Sample Flow & Pressure Sensors Status Outputs: 1 – 6 Control Inputs: 1 – 8

Pagina 96 - Passwords and Levels

Model 360E Instruction Manual THEORY OF OPERATION 10.3.2. CPU The Model 360E’s CPU is a, low power (5 VDC, 0.8A max), high performance, 386-based mic

Pagina 97

THEORY OF OPERATION Model 360E Instruction Manual 10.3.3.3. GFC Wheel A synchronous AC motor turns the GFC wheel motor. For analyzers operating on 6

Pagina 98 - Communication Ports

Model 360E Instruction Manual THEORY OF OPERATION Reference Pulses Measurement Pulses IR Beam Pulses Segment Sensor Pulses MR Sensor Pulses Figure 1

Pagina 99 - Female DB-9 (COM2)

THEORY OF OPERATION Model 360E Instruction Manual Thermo-Electric Cooler Control Circuit Sample & Hold Circuits (x4) Amplifiers Dark Switch

Pagina 100 - CN3 & CN4

Model 360E Instruction Manual THEORY OF OPERATION The four sample and hold circuits are: ACTIVE WHEN: DESIGNATION IR BEAM PASSING THROUGH SEGMENT SE

Pagina 101 - COM1 – RS-232

Model 360E Instruction Manual Getting Started 3. GETTING STARTED 3.1. Unpacking and Initial Set Up CAUTION To avoid personal injury, always use two

Pagina 102

THEORY OF OPERATION Model 360E Instruction Manual 10.3.4.4. Photo-Detector Temperature Control The synch/demod board also contains circuitry that con

Pagina 103

Model 360E Instruction Manual THEORY OF OPERATION Zero/Span Valve Options Any zero/span/shutoff valve options installed in the analyzer are controlle

Pagina 104 - 92 05232 Rev B3

THEORY OF OPERATION Model 360E Instruction Manual Figure 10-14: Location of relay board Status LED’s 10.3.5.2. I2C Watch Dog Circuitry Special circui

Pagina 105 - LED FUNCTION

Model 360E Instruction Manual THEORY OF OPERATION 10.3.6.3. Thermistor Interface This circuit provides excitation, termination and signal selection f

Pagina 106

THEORY OF OPERATION Model 360E Instruction Manual All four analog outputs are connected back to the A/D converter through a Loop-back circuit. This

Pagina 107 - PRIMARY SETUP MENU

Model 360E Instruction Manual THEORY OF OPERATION 10.3.8. Power Supply/ Circuit Breaker The analyzer operates on 100 VAC, 115 VAC or 230 VAC power at

Pagina 108

THEORY OF OPERATION Model 360E Instruction Manual 10.4. Interface The analyzer has several ways to communicate the outside world, see Figure 10-16.

Pagina 109 - 05232 Rev B3 97

Model 360E Instruction Manual THEORY OF OPERATION The most commonly used method for communicating with the M360E Analyzer is via the instrument’s fr

Pagina 110 - Administrator

THEORY OF OPERATION Model 360E Instruction Manual 10.4.1.4. Keyboard/Display Interface Electronics FRONT PANEL Keypad Decoder Key Press Dete

Pagina 111 - Multidrop

Model 360E Instruction Manual THEORY OF OPERATION I2C INTERFACE CHIP This IC performs several functions:  Using a dedicated digital status bit, it

Pagina 112 - RS-232 port

THIS PAGE IS INTENTIONALLY LEFT BLANK

Pagina 113 - 6.11.8. COM Port Baud Rate

Getting Started Model 360E Instruction Manual 5. Check the connectors of the various internal wiring harnesses and pneumatic hoses to make sure they

Pagina 114 - 6.11.9. COM Port Testing

THEORY OF OPERATION Model 360E Instruction Manual 10.5. Software Operation The Model 360E Gas Filter Correlation Carbon Dioxide Analyzer is at its he

Pagina 115 - STABIL_GAS

Model 360E Instruction Manual THEORY OF OPERATION Two conditions must be simultaneously met to switch to the short filter. First the instantaneous c

Pagina 116 - 104 05232 Rev B3

THEORY OF OPERATION Model 360E Instruction Manual 190 05232 Rev B3 10.5.5. Internal Data Acquisition System (iDAS) The iDAS is designed to implemen

Pagina 117 - Mode Indicator

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11. TROUBLESHOOTING & REPAIR PROCEDURES This contains a variety of methods

Pagina 118 - 6.13.2. Signal I/O

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.1.1. Interpreting WARNING Messages The most common and/or serious instrument

Pagina 119

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES Table 11-1: Warning Messages - Indicated Failures WARNING MESSAGE FAULT CONDI

Pagina 120

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual Table 11-1: Warning Messages – Indicated Failures (cont.) WARNING MESSAGE FAU

Pagina 121 - ANALOG OUT

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES Table 11-2: Test Functions - Indicated Failures TEST FUNCTIONS INDICATED FAI

Pagina 122

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.1.3. Using the Diagnostic Signal I/O Function The Signal I/O parameters foun

Pagina 123

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.1.4. Internal Electronic Status LED’s Several LED’s are located inside the i

Pagina 124 - ENTR accepts the new setting

Model 360E Instruction Manual Getting Started 3.1.1. M360 Layout Figure 3-1 shows the front panel layout of the analyzer. Figure 3-2 shows the rear

Pagina 125 - See Table 3-1 for

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual D1 – M/R Sensor Status D2 – Segment Sensor Status JP4 Connector to Opto-Pickup

Pagina 126

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES Table 11-5: Relay Board Status LED Failure Indications SIGNAL I/O PARAMETER L

Pagina 127

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.2. Gas Flow Problems The M360E has two main gas flow path, the sample gas fl

Pagina 128 - 6.13.4.5. AIN Calibration

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES GFC WheelHousing SAMPLE CHAMBERGFC Motor Heat SyncPurge GasFlow Contro

Pagina 129 - 6.13.5. Electric Test

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual Figure 11.9: M360E – Internal Pneumatics with O2 Sensor Option 65 11.2.2. Ty

Pagina 130 - for Measurement signal

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 4. If gas flows through the instrument when it is disconnected from its source

Pagina 131 - 6.13.7. Pressure Calibration

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.2.2.6. Sample Pump The sample pump should start immediately after the front

Pagina 132 - 6.13.8. Flow Calibration

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.3.2. Non-Repeatable Zero and Span As stated earlier, leaks both in the M360E

Pagina 133 - 6.13.9. Test Channel Output

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.4. Other Performance Problems Dynamic problems (i.e. problems which only ma

Pagina 134 - Alarms

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 4. If K2 checks out OK, the thermistor temperature sensor located on the optic

Pagina 135 - 6.15.1.1. Status Outputs

Getting Started Model 360E Instruction Manual Particulate Filter SYNC/DEMOD BOARD GFC Motor On/Off Switch & Circuit Breaker PS1 (+5 VDC; 15VD

Pagina 136 - 124 05232 Rev B3

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.4.2. Excessive Noise Noise is continuously monitored in the TEST functions a

Pagina 137 - 6.15.1.2. Control Inputs

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.5. Subsystem Checkout The preceding s of this manual discussed a variety of

Pagina 138

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual A voltmeter should be used to verify that the DC voltages are correct per the v

Pagina 139 - 6.15.2.4. Data Types

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.5.5. Relay Board The relay board PCA (04135) can be most easily checked by o

Pagina 140 - 6.15.2.5. Status Reporting

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.5.6.2. Opto Pickup Assembly Operation of the opto pickup PCA (04088) can be

Pagina 141

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 1. For Pressure related problems:  Measure the voltage across C1 it should b

Pagina 142

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual Table 11-10: Analog Output Test Function - Nominal Values Voltage Outputs F

Pagina 143

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.5.7.4. Status Outputs The procedure below can be used to test the Status out

Pagina 144 - Parameter Standard Hessen

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.5.8. CPU There are two major types of failures associated with the CPU board

Pagina 145

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.5.9.2. Troubleshooting Analyzer/Modem or Terminal Operation These are the ge

Pagina 146

Model 360E Instruction Manual Getting Started GFC Temperature Sensor Shock Absorbing Mounting Bracket Purge Gas Inlet IR Source GFC Wheel Heat Sync

Pagina 147

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11.6. Repair Procedures This contains procedures that might need to be performe

Pagina 148 - 6.15.4.8. Instrument ID Code

Model 360E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES 11.6.2. Removing/Replacing the GFC Wheel When removing or replacing the GFC Wh

Pagina 149 - 7. CALIBRATION PROCEDURES

TROUBLESHOOTING & REPAIR PROCEDURES Model 360E Instruction Manual 11 12 Figure 11-12: Removing the GFC Wheel 12. Remove the GFC wheel/mask asse

Pagina 150 - VENT here if input

Model 360E Instruction Manual A Primer on Electro-Static Discharge 12. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne Instruments considers the preve

Pagina 151

A Primer on Electro-Static Discharge Model 360E Instruction Manual 12.2. How Electro-Static Charges Cause Damage Damage to components occurs when the

Pagina 152

Model 360E Instruction Manual A Primer on Electro-Static Discharge 12.3. Common Myths About ESD Damage  I didn’t feel a shock so there was no elec

Pagina 153

A Primer on Electro-Static Discharge Model 360E Instruction Manual For technicians that work in the field, special lightweight and portable anti-ESD

Pagina 154

Model 360E Instruction Manual A Primer on Electro-Static Discharge 12.4.2. Basic anti-ESD Procedures for Analyzer Repair and Maintenance 12.4.2.1. W

Pagina 155

A Primer on Electro-Static Discharge Model 360E Instruction Manual 12.4.2.3. Transferring Components from Rack to Bench and Back When transferring a

Pagina 156

Model 360E Instruction Manual A Primer on Electro-Static Discharge 12.4.2.5. Packing Components for Return to Teledyne Instruments Customer Service.

Pagina 157

Getting Started Model 360E Instruction Manual 3.1.2. Electrical Connections 3.1.2.1. Power Connection Attach the power cord to the analyzer and plug

Pagina 158

A Primer on Electro-Static Discharge Model 360E Instruction Manual 228 05232 Rev B3 USER NOTES:

Pagina 159 - Standard Procedure

Model 360E Instruction Manual APPENDIX A - Version Specific Software Documentation APPENDIX A - Version Specific Software Documentation APPENDIX A-1:

Pagina 160

APPENDIX A-1: M360E Software Menu Trees, Revision G.4 Model 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 CO25 O25 Z

Pagina 161

Model 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 SAMPLE <TST TST> SETUP COMM DIAG VARS (Secondary Setup Me

Pagina 162 - 150 05232 Rev B3

APPENDIX A-1: M360E Software Menu Trees, Revision G.4 Model 360E Instruction Manual • MODEL NAME • SERIAL NUMBER • SOFTWARE REVISION • LIBRARY R

Pagina 163 - 7.7. Calibration Quality

Model 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 NAME EVENT PARAMETERS REPORT PERIOD NUMBER OF RECORDS RS-232 REP

Pagina 164

APPENDIX A-1: M360E Software Menu Trees, Revision G.4 Model 360E Instruction Manual CFG ACAL1 SETUP CLK PASS RNGE DAS MORE COMM VARS DIAG Go To DI

Pagina 165 - 8. EPA PROTOCOL CALIBRATION

Model 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 INSTRUMENT IP5 GATEWAY IP5 SUBNET MASK5 TCP PORT3 DHCP INSTRUME

Pagina 166

APPENDIX A-1: M360E Software Menu Trees, Revision G.4 Model 360E Instruction Manual . See Fig A-5 QUIET COMPUTER SECURITY HESSEN PROTOCOL COM[1,2] E

Pagina 167 - 9.1. Maintenance Schedule

odel 360E Instruction Manual APPENDIX A-1: M360E Software Menu Trees, Revision G.4 3 Rev G.4 A-9 CFG ACAL1 CLK PASS RNGE DAS MORE COMM VARS ANALOG

Pagina 168

Model 360E Instruction Manual Getting Started 3.1.2.2. Output Connections The M360E is equipped with several analog output channels accessible throug

Pagina 169 - Warm-up

APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 T

Pagina 170

Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 171 - 9.3. Maintenance Procedures

APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 172 - 9.3.3.2. Pressure Leak Check

Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 173

APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 174

Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 175 - 10.1. Measurement Method

APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 176 - 10.1.3.1. The GFC Wheel

Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 177

APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 178 - Chamber

Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 179 - Interference Rejection

Getting Started Model 360E Instruction Manual 3.1.2.3. Connecting the Status Outputs If you wish utilize the analyzer’s status outputs to interface w

Pagina 180 - 10.2. Pneumatic Operation

APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 181 - 10.2.2. Flow Rate Control

Model 360E Instruction Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 182 - 10.2.5. Pneumatic Sensors

APPENDIX A-2: Setup Variables For Serial I/O, Revision G.4 Model 360E Instruction Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VALUE VALUE RANGE DESC

Pagina 183 - 10.3. Electronic Operation

Model 360E Instruction Manual APPENDIX A-3: Warnings and Test Functions, Revision G.4 APPENDIX A-3: Warnings and Test Functions, Revision G.4 Table A

Pagina 184 - PC 104 Bus

APPENDIX A-3: Warnings and Test Functions, Revision G.4 Model 360E Instruction Manual Table A-3: M360E Test Functions, Revision G.4 TEST FUNCTION

Pagina 185 - 10.3.2. CPU

Model 360E Instruction Manual APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4 APPENDIX A-4: Signal I/O Definitions f

Pagina 186 - 10.3.3.3. GFC Wheel

APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4 Model 360E Instruction Manual SIGNAL NAME BIT OR CHANNEL NUMBER DESCRI

Pagina 187 - 10.3.4.1. Overview

Model 360E Instruction Manual APPENDIX A-4: Signal I/O Definitions for 300E Series Analyzers, Revision G.4 SIGNAL NAME BIT OR CHANNEL NUMBER DESCRI

Pagina 188

APPENDIX A-5: M360E iDAS Parameters Revision G.4 Model 360E Instruction Manual APPENDIX A-5: M360E iDAS Parameters Revision G.4 Table A-5: M360E

Pagina 189 - D2 Segment Sensor

Model 360E Instruction Manual APPENDIX A-5: M360E iDAS Parameters Revision G.4 Table A-6: M360E iDAS Functions, Revision G.4 NAME DESCRIPTION U

Pagina 190 - 10.3.5. Relay Board

Model 360E Instruction Manual Getting Started 3.1.2.4. Connecting the Control Inputs If you wish to use the analyzer to remotely activate the zero an

Pagina 191 - PN 04135

APPENDIX A-6: Terminal Command Designators, Revision G.4 Model 360E Instruction Manual APPENDIX A-6: Terminal Command Designators, Revision G.4 Table

Pagina 192 - C Watch Dog Circuitry

Model 360E Instruction Manual APPENDIX A-6: Terminal Command Designators, Revision G.4 Table A-8: Terminal Key Assignments, Revision G.4 TERMINAL

Pagina 193 - 10.3.6.4. Analog Outputs

APPENDIX A-6: Terminal Command Designators, Revision G.4 Model 360E Instruction Manual USER NOTES: A-32 05233 Rev G.4

Pagina 194 - C Data Bus

Model 360E Instruction Manual APPENDIX B - M360E Spare Parts List 05234 Rev B B-1 APPENDIX B - M360E Spare Parts List NOTE Use of replacement parts

Pagina 195

APPENDIX B - M360E Spare Parts List Model 360E Instruction Manual B-2 05234 Rev B

Pagina 196 - 10.4. Interface

M360E Spare Parts ListPart NumberDescription000941000 ORIFICE, 13 MIL (SAMPLE FLOW)001760400 ASSY, FLOW CTL, 800CC, 1/

Pagina 197 - 10.4.1.3. Display

M360E Spare Parts ListPart NumberDescription050320000 PCA, OPTO-INTERRUPTER, M300E 052560000 PCA, OPTO-INTE

Pagina 198 - C to/from CPU

Recommended Spare Parts Stocking Levels Model 360E1 2-5 6-10 11-20 21-30003290500 ASSY, THERMISTOR, M300 BENCH 1 1 1 1 200

Pagina 199

THIS PAGE IS INTENTIONALLY LEFT BLANK

Pagina 200 - 10.5. Software Operation

Model 360E Instruction Manual Warranty/Repair Questionnaire TELEDYNE INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies C

Pagina 201

Getting Started Model 360E Instruction Manual 3.1.2.5. Connecting the Serial Ports If you wish to utilize either of the analyzer’s two serial interfa

Pagina 202

Model 360E Instruction Manual Warranty/Repair Questionnaire TELEDYNE INSTRUMENTS Advanced Pollution Instrumentation A Teledyne Technologies C

Pagina 203

Model360E Instruction Manual APPENDIX D - ELECTRONIC SCHEMATICS 05236 Rev C D-1 APPENDIX D - ELECTRONIC SCHEMATICS Table D-1: List of Included

Pagina 204

APPENDIX D - ELECTRONIC SCHEMATICS Model 360E Instruction Manual D-2 05236 Rev C

Pagina 205

Model 360E Instruction Manual Getting Started Table 3-4: Model 360E Rear Panel Pneumatic Connections REAR PANEL LABEL FUNCTION SAMPLE Connect a

Pagina 206

Model 360E Instruction Manual M360E Documentation SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many imp

Pagina 207 - 05232 Rev B3 195

Getting Started Model 360E Instruction Manual VENT Figure 3-7: Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator 1. Attac

Pagina 208

Model 360E Instruction Manual Getting Started EXAMPLE: If the application is to measure between 0 ppm and 500 ppm, an appropriate Span Gas would be

Pagina 209 - P/N 04069

Getting Started Model 360E Instruction Manual 3.1.3.2. Connections with Internal Valve Options Installed Figures 3-8 through 3-11 show the proper pne

Pagina 210

Model 360E Instruction Manual Getting Started Figure 3-10: Pneumatic Connections–M360E with Zero/Span Valves (OPT 52) Source ofSAMPLE GASRemove

Pagina 211 - 05232 Rev B3 199

Getting Started Model 360E Instruction Manual 3.1.3.3. Pneumatic Connections to M360E in Multipoint Calibration Applications Some applications may re

Pagina 212 - 11.2. Gas Flow Problems

Model 360E Instruction Manual Getting Started 3.2. Initial Operation If you are unfamiliar with the M360E theory of operation, we recommend that you

Pagina 213 - SAMPLE CHAMBER

Getting Started Model 360E Instruction Manual 3.2.2. Warm Up The M360E requires about 30 minutes warm-up time before reliable CO2 measurements can be

Pagina 214 - 11.2.2.1. Flow is Zero

Model 360E Instruction Manual Getting Started Table 3-6: Possible Warning Messages at Start-Up MESSAGE MEANING ANALOG CAL WARNING The instrument’

Pagina 215 - 11.2.2.3. High Flow

Getting Started Model 360E Instruction Manual 3.2.4. Functional Check 1. After the analyzer’s components has warmed up for at least 30 minutes, veri

Pagina 216 - 11.3. Calibration Problems

Model 360E Instruction Manual Getting Started 3.3. Initial Calibration Procedure The next task is to calibrate the analyzer. To perform the followi

Pagina 218 - 11.4.1.2. Bench Temperature

Getting Started Model 360E Instruction Manual 2. Set the expected CO2 span gas concentration This sequence causes the analyzer to prompt for the e

Pagina 219

Model 360E Instruction Manual Getting Started 3. Perform the Zero/Span Calibration Procedure M-P CAL STABIL=XXX.X PPM CO2 =XXX

Pagina 220 - 11.4.2. Excessive Noise

Getting Started Model 360E Instruction Manual 3.3.1. Initial O2 Sensor Calibration Procedure If your instrument has an O2 sensor option installed tha

Pagina 221 - 11.5. Subsystem Checkout

Model 360E Instruction Manual Getting Started STEP 2 – ACTIVATE O2 SENSOR STABILITY FUNCTION To change the stability test function from NOx concentra

Pagina 222 - 11.5.3. I

Getting Started Model 360E Instruction Manual STEP 4 – O2 Zero/Span Calibration: To perform the zero/span calibration procedure: 32 05232 Rev B3

Pagina 223 - 11.5.6. Sensor Assembly

Model 360E Instruction Manual Getting Started 05232 Rev B3 33 The Model 360E Analyzer is now ready for operation NOTE Once you have completed the a

Pagina 225 - 11.5.7. Motherboard

Model 360E Instruction Manual Frequently Asked Questions 4. FREQUENTLY ASKED QUESTIONS 4.1. FAQ’s The following is a list from the Teledyne Instrumen

Pagina 226

Frequently Asked Questions Model 360E Instruction Manual A: These components are located inside the flow control assembly that is attached to the inl

Pagina 227 - 11.5.7.4. Status Outputs

Model 360E Instruction Manual Frequently Asked Questions 05232 Rev B3 37 IC – Acronym for Integrated Circuit, a modern, semi-conductor circuit that

Pagina 228 - 11.5.8. CPU

Model 360E Instruction Manual M360E Documentation TABLE OF CONTENTS 1. M360E DOCUMENTATION...

Pagina 230 - 11.6. Repair Procedures

Model 360E Instruction Manual Optional Hardware and Software 5. OPTIONAL HARDWARE AND SOFTWARE This includes a brief description of the hardware and

Pagina 231

Optional Hardware and Software Model 360E Instruction Manual 5.2.1. Converting Current Loop Analog Outputs to Standard Voltage Outputs. NOTE Servicin

Pagina 232

Model 360E Instruction Manual Optional Hardware and Software 5.4. Calibration Valves Options There are four available options involving Zero/Span/Shu

Pagina 233 - 10-25% RH

Optional Hardware and Software Model 360E Instruction Manual GFC WheelHousing SAMPLE CHAMBERGFC Motor Heat SyncPurge GasFlow ControlExternal

Pagina 234

Model 360E Instruction Manual Optional Hardware and Software Table 5-2: Zero/Span Valve Operating States for Options 52 & 53 Mode Valve Cond

Pagina 235 - Ground Point

Optional Hardware and Software Model 360E Instruction Manual 5.4.4. Zero/Span Valve with External CO2 Scrubber (Option 53) Option 53 is operationally

Pagina 236

Model 360E Instruction Manual Optional Hardware and Software 5.5.3. Ethernet (Option 63) When installed, this option is electronically connected to t

Pagina 237

Optional Hardware and Software Model 360E Instruction Manual 5.6. Oxygen Sensor (OPT 65) 5.6.1. Theory of Operation 5.6.1.1. Paramagnetic measurement

Pagina 238

Model 360E Instruction Manual Optional Hardware and Software The O2 concentration range is 0-100% (user selectable) with 0.1% precision and accuracy

Pagina 239 - User Notes:

M360E Documentation Model 360E Instruction Manual 6. OPERATING INSTRUCTIONS...

Pagina 240 - USER NOTES:

Optional Hardware and Software Model 360E Instruction Manual 5.7. Additional Manuals 5.7.1. Printed Manuals (Option 70) Additional printed copies of

Pagina 241

Model 360E Instruction Manual Optional Hardware and Software 05232 Rev B3 49 5.9.2. Maintenance Mode Switch API’s instruments can be equipped with a

Pagina 243 - 05233 Rev G.4 A-3

Model 360E Instruction Manual Operating Instructions 6. OPERATING INSTRUCTIONS To assist in navigating the analyzer’s software, a series of menu tree

Pagina 244 - A-4 05233 Rev G.4

Operating Instructions Model 360E Instruction Manual Table 6-1: Analyzer Operating modes MODE MEANING DIAG One of the analyzer’s diagnostic modes

Pagina 245 - 05233 Rev G.4 A-5

Model 360E Instruction Manual Operating Instructions Table 6-2: Test Functions Defined PARAMETER DISPLAY TITLE UNITS MEANING Range RANGE RANGE11

Pagina 246 - A-6 05233 Rev G.4

Operating Instructions Model 360E Instruction Manual RANGERANGE11 RANGE21 02 RANGE STABIL MEAS REF MR RATIO PRES SAMP FL SAMP TEMP BENCH TEMPWHEEL T

Pagina 247 - 05233 Rev G.4 A-7

Model 360E Instruction Manual Operating Instructions 6.2.2. Warning Messages The most common instrument failures will be reported as a warning on the

Pagina 248 - A-8 05233 Rev G.4

Operating Instructions Model 360E Instruction Manual To view and clear warning messages SAMPLE HVPS WARNING CO2 = 0.00

Pagina 249 - 3 Rev G.4 A-9

Model 360E Instruction Manual Operating Instructions 6.4. SETUP MODE The SETUP mode contains a variety of choices that are used to configure the anal

Pagina 250 - A-10 05233 Rev G.4

Model 360E Instruction Manual M360E Documentation 6.13.8. Flow Calibration...

Pagina 251 - 05233 Rev G.4 A-11

Operating Instructions Model 360E Instruction Manual 6.5. SETUP  CFG: Viewing the Analyzer’s Configuration Information Pressing the CFG key displays

Pagina 252 - A-12 05233 Rev G.4

Model 360E Instruction Manual Operating Instructions 6.7. SETUP  DAS: Using the Data Acquisition System (iDAS) The M360E analyzer contains a flexibl

Pagina 253 - 05233 Rev G.4 A-13

Operating Instructions Model 360E Instruction Manual 6.7.1.1. iDAS Channels The key to the flexibility of the iDAS is its ability to store a large nu

Pagina 254 - A-14 05233 Rev G.4

Model 360E Instruction Manual Operating Instructions 6.7.1.2. iDAS Parameters Data parameters are types of data that may be measured and stored by th

Pagina 255 - 05233 Rev G.4 A-15

Operating Instructions Model 360E Instruction Manual 6.7.1.3. iDAS Triggering Events Triggering events define when and how the iDAS records a measure

Pagina 256 - A-16 05233 Rev G.4

Model 360E Instruction Manual Operating Instructions Note The CALDAT, STBZRO and STBSPN channels collect data based on events (e.g. a calibration op

Pagina 257 - 05233 Rev G.4 A-17

Operating Instructions Model 360E Instruction Manual NOTE Sending an iDAS configuration to the analyzer through its COM ports will replace the existi

Pagina 258 - A-18 05233 Rev G.4

Model 360E Instruction Manual Operating Instructions 6.7.2.1. Viewing iDAS Data and Settings iDAS data and settings can be viewed on the front panel

Pagina 259 - 05233 Rev G.4 A-19

Operating Instructions Model 360E Instruction Manual 6.7.2.2. Editing iDAS Data Channels iDAS configuration is most conveniently done through the API

Pagina 260 - A-20 05233 Rev G.4

Model 360E Instruction Manual Operating Instructions From the end of the previous key sequence … SETUP X.X NAME:CONC C O

Pagina 261 - 05233 Rev G.4 A-21

M360E Documentation Model 360E Instruction Manual 10.2.1. Sample Gas Flow ...

Pagina 262 - A-22 05233 Rev G.4

Operating Instructions Model 360E Instruction Manual 6.7.2.4. Editing iDAS Parameters Data channels can be edited individually from the front panel w

Pagina 263

Model 360E Instruction Manual Operating Instructions To configure a specific data parameter, press: FROM THE EDIT DATA PARAMETER MENU (see previo

Pagina 264

Operating Instructions Model 360E Instruction Manual 6.7.2.5. Sample Period and Report Period The iDAS defines two principal time periods by which sa

Pagina 265 - 05233 Rev G.4 A-25

Model 360E Instruction Manual Operating Instructions To define the REPORT PERIOD, follow the instruction shown in Section 6.7.2.2 then press: Edit D

Pagina 266 - A-26 05233 Rev G.4

Operating Instructions Model 360E Instruction Manual 6.7.2.6. Number of Records The number of data records in the M360E is limited to about a cumulat

Pagina 267 - 05233 Rev G.4 A-27

Model 360E Instruction Manual Operating Instructions 6.7.2.7. RS-232 Report Function The M360E iDAS can automatically report data to the communicatio

Pagina 268 - A-28 05233 Rev G.4

Operating Instructions Model 360E Instruction Manual 6.7.2.10. Disabling/Enabling Data Channels Data channels can be temporarily disabled, which can

Pagina 269 - 05233 Rev G.4 A-29

Model 360E Instruction Manual Operating Instructions 6.7.3. Remote iDAS Configuration Editing channels, parameters and triggering events as described

Pagina 270 - A-30 05233 Rev G.4

Operating Instructions Model 360E Instruction Manual Figure 6-6: iDAS Configuration Through a Terminal Emulation Program. Both procedures are be

Pagina 271

Model 360E Instruction Manual Operating Instructions 6.8. SETUP  RNGE: Analog Output Reporting Range Configuration The analyzer has three active ana

Pagina 272

Model 360E Instruction Manual M360E Documentation 11.2.3. Poor or Stopped Flow of Purge Gas...

Pagina 273 - 05234 Rev B B-1

Operating Instructions Model 360E Instruction Manual The M360E solves this problem by allowing the user to select a scaled reporting range for the an

Pagina 274

Model 360E Instruction Manual Operating Instructions NOTE Upper span limit setting for the individual range modes are shared. Resetting the span lim

Pagina 275 - M360E Spare Parts List

Operating Instructions Model 360E Instruction Manual 6.8.4. Dual Range Mode (DUAL) Selecting Dual Range mode allows the A1 and A2 outputs to be confi

Pagina 276

Model 360E Instruction Manual Operating Instructions 6.8.5. Auto Range Mode (AUTO) In AUTO range mode, the analyzer automatically switches the report

Pagina 277 - Model 360E

Operating Instructions Model 360E Instruction Manual 6.8.6. Range Units The M360E can display concentrations in parts per billion (109 mols per mol,

Pagina 278

Model 360E Instruction Manual Operating Instructions 6.8.7. Dilution Ratio The dilution ratio is a software option that allows the user to compensate

Pagina 279 - Questionnaire

Operating Instructions Model 360E Instruction Manual 6.9. SETUP  PASS: Password Feature The M360E provides password protection of the calibration an

Pagina 280

Model 360E Instruction Manual Operating Instructions 6.10. SETUP  CLK: Setting the Internal Time-of-Day Clock The M360E has a time of day clock tha

Pagina 281

Operating Instructions Model 360E Instruction Manual 6.11. SETUP  MORE COMM: Using the Analyser’s Communication Ports The M360E is equipped with tw

Pagina 282 - D-2 05236 Rev C

Model 360E Instruction Manual Operating Instructions 6.11.2. COMM Port Default Settings Received from the factory, the analyzer is set up to emulate