Teledyne 360E Manual de usuario Pagina 66

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 282
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 65
Operating Instructions Model 360E Instruction Manual
RANGE
RANGE1
1
RANGE2
1
0
2
RANGE
STABIL
MEAS
REF
MR RATIO
PRES
SAMP FL
SAMP TEMP
BENCH TEMP
WHEEL TEMP
BOX TEMP
PHT DRIVE
SLOPE
OFFSET
TEST
TIME
SAMPLE RANGE = 500.000 PPM CO2 = XXX.X
< TST TST > CAL SETUP
1
Only appears instrument is set
for DUAL or AUTO reporting
range modes.
2
Only appears if 0
2
Sensor
Option is installed.
Refer to
Table 6-2 for
definitions of
these test
functions.
Toggle <TST TST> keys to
scroll throu
g
h list of functions
Figure 6-3: Viewing M360E TEST Functions
NOTE
A value of “XXXX” displayed for any of the TEST functions indicates an out-of-range
reading or the analyzer’s inability to calculate it.
All pressure measurements are represented in terms of absolute pressure. Absolute,
atmospheric pressure is 29.92 in-Hg-A at sea level. It decreases about 1 in-Hg per 300
m gain in altitude. A variety of factors such as air conditioning and passing storms can
cause changes in the absolute atmospheric pressure.
54 05232 Rev B3
Vista de pagina 65
1 2 ... 61 62 63 64 65 66 67 68 69 70 71 ... 281 282

Comentarios a estos manuales

Sin comentarios