Analyzer Instruction Manual, 02246, Rev. G, DCN 5247
1. In SAMPLE mode, scroll the TEST functions to SAMP FLW. Observe the reading. Then
subtract the observed reading from the desired reading.
2.
Go to the VARS menu by SETUP-MORE-VARS, then pressing ENTR. Scroll to
SFLOW_SET and press EDIT. The first value displayed is the SFLOW_SET:VALUE
reading, ADD the value from step 1 to the reading shown and key in the new reading, then
press ENTR. Check the low and high warning limits which are displayed as the next 2 values
to make sure the new value does not exceed the warning limits. Press exit to return to
SAMPLE mode.
3.
Observe the SAMP FLW value. If necessary repeat step 2 to adjust the reading again to
match the desired flow rate.
9-58
PRINTED DOCUMENTS ARE UNCONTROLLED