Teledyne API Model 300M CO Analyzer Instruction Manual, 04033, Rev. A
Examples of possible sequences are as following under any one of three available SEQx.
Example 1: To perform a 15 minute zero check once per day at 10:30 PM, 12/20/93.
1. MODE: ZERO
2. TIMER ENABLED: ON
3. STARTING DATE: 12/20/93
4. STARTING TIME: 22:30
5. DELTA DAYS: 1
6. DELTA TIME: 00:00
7. DURATION: 15
8. CALIBRATE: OFF
Example 2: To perform a 15 min zero span check once per day retarding 15 minutes everyday
starting at 11:30 PM, 12/20/93.
1. MODE: ZERO-SPAN
2. TIMER ENABLED: ON
3. STARTING DATE: 12/20/93
4. STARTING TIME: 23:30
5. DELTA DAYS: 0
6. DELTA TIME: 23:45
7. DURATION: 15
8. CALIBRATE: OFF
Example 3: To perform span check once per week starting at 11:30 PM, 12/20/93.
1. MODE: SPAN
2. TIMER ENABLED: ON
3. STARTING DATE: 12/20/93
4. STARTING TIME: 23:30
5. DELTA DAYS: 7
6. DELTA TIME: 00:00
7. DURATION: 15
8. CALIBRATE: OFF
3-7