Teledyne 300EU Manual de usuario Pagina 63

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 94
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 62
M300EU THEORY OF OPERATION
(Addendum to M300EU Manual - P/N 04145)
10.3.2.2. Oven Heaters AC Power Configuration
The two main heaters for the convection oven area of the M300EU are AC heaters. A jumper, located in line on
the power connection between the relay PCA and the heaters, set the heaters in one of two configurations:
In parallel for 115 VAC operation, or;
In series for 230 VAC operation.
Blue Jumper
115 VAC Configuration
White Jumpers
230 VAC Configuration
Figure 10-6: M300EU Oven Heater Configuration Jumpers
Figure 10-7: M300EU Oven Heater Configuration Circuit
05514 Rev A1 55
Vista de pagina 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 93 94

Comentarios a estos manuales

Sin comentarios